POTENTIOMETRIC DETECTION IN FLOW-INJECTION WITHOUT USE OF A CONVENTIONAL REFERENCE ELECTRODE

被引:12
作者
LIU, RM
LIU, DJ
SUN, AL
机构
[1] Department of Chemistry, Liaocheng Teachers College, Liaocheng, Shandong
关键词
ION-SELECTIVE ELECTRODE; POTENTIOMETRIC DETECTION; FLOW INJECTION;
D O I
10.1039/an9921701335
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An ion-selective electrode-flow injection system has been developed with two ion-selective electrodes arranged in parallel in a flow injection system. The system does not contain a conventional reference electrode,and no liquid junction is present. The sample is injected into the two carrier streams alternately and gives a potential response at the electrode. The proposed method is simple and convenient and was applied successfully to the determination of fluoride and nitrate in natural waters, the determination of sodium and potassium in natural waters and the determination of potassium and nitrate in soil extracts.
引用
收藏
页码:1335 / 1337
页数:3
相关论文
共 11 条
[1]   FLOW-INJECTION ANALYSIS WITH ELECTROCHEMICAL DETECTION [J].
CAMMANN, K .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1988, 329 (06) :691-697
[2]   SIMULTANEOUS DETERMINATIONS IN FLOW-INJECTION ANALYSIS - A REVIEW [J].
DECASTRO, MDL ;
CASES, MV .
ANALYST, 1984, 109 (04) :413-419
[3]   SIMULTANEOUS DETERMINATION OF TRACE AMOUNTS OF IRON(III) AND TITANIUM(IV) BY FLOW-INJECTION WITH SPECTROPHOTOMETRIC DETECTION [J].
KOZUKA, S ;
SAITO, K ;
OGUMA, K ;
KURODA, R .
ANALYST, 1990, 115 (04) :431-434
[5]  
OGUMA K, 1985, J FLOW INJECT ANAL, V2, P98
[6]   NEW APPROACH TO THE SIMULTANEOUS DETERMINATION OF POLLUTANTS IN WASTE-WATERS BY FLOW-INJECTION ANALYSIS .1. ANIONIC POLLUTANTS [J].
RIOS, A ;
DECASTRO, MDL ;
VALCARCEL, M .
ANALYST, 1984, 109 (11) :1487-1492
[7]  
WANG F, 1982, LIZI XUANZE DIANJI T, V2, P46
[8]  
XIE S, 1985, ANAL TECHNIQUE ION S, P344
[9]  
YU Q, 1981, LIZI XUANZE DIANJI T, V1, P50
[10]  
ZHANG J, 1981, LIZI XUANZE DIANJI T, V1, P50