AN AUTOMATED THERMAL RELAXATION CALORIMETER FOR OPERATION AT LOW-TEMPERATURE (0.5K-LESS-THAN-T-LESS-THAN-10K)

被引:2
作者
BANERJEE, S
PRINS, MWJ
RAJEEV, KP
RAYCHAUDHURI, AK
机构
[1] Department of Physics, Indian Institute of Science, Bangalore
来源
PRAMANA-JOURNAL OF PHYSICS | 1992年 / 39卷 / 04期
关键词
RELAXATION CALORIMETER; SPECIFIC HEAT; LOW TEMPERATURE;
D O I
10.1007/BF02845823
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We describe an automated calorimeter for measurement of specific heat in the temperature range 10 K > T > 0.5 K. It uses sample of moderate size (100-1000 mg), has a moderate precision and accuracy (2%-5%), is easy to operate and the measurements can be done quickly with He-4 economy. The accuracy of this calorimeter was checked by measurement of specific heat of copper and that of aluminium near its superconducting transition temperature.
引用
收藏
页码:391 / 404
页数:14
相关论文
共 11 条
[1]   SELECTION OF A THERMAL BONDING AGENT FOR TEMPERATURES BELOW 1K [J].
ANDERSON, AC ;
PETERSON, RE .
CRYOGENICS, 1970, 10 (05) :430-&
[2]   HEAT-CAPACITY MEASUREMENTS ON SMALL SAMPLES AT LOW-TEMPERATURES [J].
BACHMANN, R ;
SCHWALL, RE ;
THOMAS, HU ;
ZUBECK, RB ;
KING, CN ;
KIRSCH, HC ;
DISALVO, FJ ;
GEBALLE, TH ;
LEE, KN ;
HOWARD, RE ;
GREENE, RL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1972, 43 (02) :205-&
[3]   LOW-TEMPERATURE CALORIMETRY FOR SMALL SAMPLES [J].
DEPUYDT, JM ;
DAHLBERG, ED .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1986, 57 (03) :483-486
[4]   SPECIFIC-HEAT MEASUREMENTS IN SMALL SAMPLES AT LOW-TEMPERATURE [J].
DUTZI, J ;
PATTALWAR, SM ;
DIXIT, RN ;
SHETE, SY .
PRAMANA, 1988, 31 (04) :253-262
[5]  
LAKSHMIKUMAR ST, 1981, RECENT DEV TECHNIQUE
[6]   HEAT CAPACITY OF ALUMINUM BETWEEN 0.1-DEGREES-K AND 4.0-DEGREES-K [J].
PHILLIPS, NE .
PHYSICAL REVIEW, 1959, 114 (03) :676-685
[7]  
RAYCHAUDHURI A, 1980, THESIS CORNELL U
[8]   FULLY AUTOMATED CALORIMETER FOR SMALL SAMPLES WITH IMPROVED SENSITIVITY [J].
REGELSBERGER, M ;
WERNHARDT, R ;
ROSENBERG, M .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1986, 19 (07) :525-532
[9]   THERMAL RELAXATION CALORIMETRY BELOW 1 K [J].
SCHUTZ, RJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1974, 45 (04) :548-551
[10]   ANALYSIS OF THE LUMPED TAU-2 EFFECT IN RELAXATION CALORIMETRY [J].
SHEPHERD, JP .