NEUTRAL CARRIER SODIUM ION-SELECTIVE MICRO-ELECTRODE FOR INTRACELLULAR STUDIES

被引:257
作者
STEINER, RA [1 ]
OEHME, M [1 ]
AMMANN, D [1 ]
SIMON, W [1 ]
机构
[1] SWISS FED INST TECHNOL,DEPT ORGAN CHEM,UNIV STR 16,CH-8092 ZURICH,SWITZERLAND
关键词
D O I
10.1021/ac50039a008
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Na+-selective microelectrodes based on new, synthetic, electrically neutral ion carriers are described. The selectivities in respect to K+ and Mg2+ are adequate for intracellular measurements of Na+ ion activities≥10 mM at ionic backgrounds of 100 mM K+, 2 mM Mg2+, 10-7 M Ca2+. Electrodes with tip diameters of around 1 μm have an electrical resistance of about 1010Ω. and a 90% response time as specified by the IUPAC recommendation of ≤5 s. © 1979, American Chemical Society. All rights reserved.