CALIBRATION OF A PROMPT GAMMA IN-VIVO NEUTRON-ACTIVATION FACILITY FOR DIRECT MEASUREMENT OF TOTAL-BODY PROTEIN IN INTENSIVE-CARE PATIENTS

被引:19
作者
MITRA, S [1 ]
PLANK, LD [1 ]
HILL, GL [1 ]
机构
[1] UNIV AUCKLAND,AUCKLAND HOSP,DEPT SURG,AUCKLAND,NEW ZEALAND
关键词
D O I
10.1088/0031-9155/38/12/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Auckland in vivo prompt gamma neutron activation analysis facility has been calibrated for direct measurement of protein and chlorine in intensive care patients without reliance on hydrogen as an internal standard. These patients often present with massive fluid overload and the technique using hydrogen as an internal standard may be invalidated. Eight variable anthropomorphic phantoms containing physiological amounts of the major body elements were constructed from commercially available flexible polythene tubes and used for calibration. These phantoms are easily disassembled and reassembled to a different size and shape. Accuracy and precision for protein and chlorine were assessed by repeated scanning of a 64 kg anthropomorphic minced-meat phantom followed by chemical analysis of the meat. Reproducibilities for protein and chlorine were 2.9% and 10.6%, respectively, for a mean dose equivalent to a subject of approximately 0.15 mSv. Chemical analysis for protein and chlorine did not differ significantly from the scanning results.
引用
收藏
页码:1971 / 1975
页数:5
相关论文
共 11 条
[1]   A PROMPT GAMMA INVIVO NEUTRON-ACTIVATION ANALYSIS FACILITY FOR MEASUREMENT OF TOTAL-BODY NITROGEN IN THE CRITICALLY ILL [J].
BEDDOE, AH ;
ZUIDMEER, H ;
HILL, GL .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (04) :371-383
[2]  
DILMANIAN FA, 1990, IN VIVO BODY COMPOSI, P309
[3]   IMPROVED CLINICAL FACILITY FOR INVIVO NITROGEN MEASUREMENT [J].
KRISHNAN, SS ;
MCNEILL, KG ;
MERNAGH, JR ;
BAYLEY, AJ ;
HARRISON, JE .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (04) :489-499
[4]   INVIVO MEASUREMENT OF TOTAL-BODY CHLORINE USING THE 8.57 MEV PROMPT DEEXCITATION FOLLOWING THERMAL-NEUTRON CAPTURE [J].
MITRA, S ;
PLANK, LD ;
KNIGHT, GS ;
HILL, GL .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (01) :161-172
[5]   VERSATILE VARIABLE ANTHROPOMORPHIC PHANTOM [J].
OXBY, CB ;
BROOKS, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (02) :440-442
[6]  
RYDE SJS, 1990, ADV IN VIVO BODY COM, P347
[7]  
STROUD DB, 1990, ADV IN VIVO BODY COM, P177
[8]   INVIVO MEASUREMENT OF TOTAL-BODY CARBON USING PU-238/BE NEUTRON SOURCES [J].
SUTCLIFFE, JF ;
MITRA, S ;
HILL, GL .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (08) :1089-1098
[9]   USE OF BODY HYDROGEN AS AN INTERNAL STANDARD IN THE MEASUREMENT OF NITROGEN INVIVO BY PROMPT NEUTRON-CAPTURE GAMMA-RAY ANALYSIS [J].
VARTSKY, D ;
PRESTWICH, WV ;
THOMAS, BJ ;
DABEK, JT ;
CHETTLE, DR ;
FREMLIN, JH ;
STAMMERS, K .
JOURNAL OF RADIOANALYTICAL CHEMISTRY, 1979, 48 (1-2) :243-252
[10]  
1971, NCRP38 REP