MASSIVE HEMOPTYSIS SECONDARY TO FLOW-DIRECTED THERMODILUTION CATHETERS

被引:23
作者
HAAPANIEMI, J
GADOWSKI, R
NAINI, M
GREEN, H
MACKENZIE, D
RUBENFIRE, M
机构
[1] SINAI HOSP,DEPT MED,CARDIOVASC DIS SECT,DETROIT,MI 48235
[2] WAYNE STATE UNIV,DETROIT,MI 48202
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1979年 / 5卷 / 02期
关键词
cardiac output; hemoptysis; monitoring; pulmonary complication; Swan‐Ganz; thermodilution;
D O I
10.1002/ccd.1810050209
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Hemoptysis is an unusual complication of flow‐directed (Swan‐Ganz) catheters. Over‐inflation of the balloon with a shearing‐induced rupture of a small pulmonary artery, and the spear effect of the catheter tip appear to be the mechanisms in the two cases presented. Diligent care to avoid overinflation of the balloon in the pulmonary capillary wedge position by observation of the pressure waveform is critical. The spear effect that is frequently seen during insertion may be eliminated by deflating the balloon at the first appearance of the pulmonary artery waveform and gradual advancement of the catheter five to eight cm, when the balloon is then reinflated to obtain the wedge. Copyright © 1979 John Wiley & Sons, Ltd.
引用
收藏
页码:151 / 157
页数:7
相关论文
共 11 条
[1]  
CHUN GMH, 1971, NEW ENGL J MED, V284, P1041
[2]  
COLVIN MP, 1975, BRIT J ANAESTH, V47, P1107
[3]   PULMONARY COMPLICATIONS OF FLOW-DIRECTED BALLOON-TIPPED CATHETER [J].
FOOTE, GA ;
SCHABEL, SI ;
HODGES, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1974, 290 (17) :927-931
[4]   BEDSIDE DIAGNOSIS OF LATENT CARDIAC COMPLICATIONS IN ACUTELY ILL PATIENTS [J].
FORRESTER, JS ;
SWAN, HJC ;
DIAMOND, GA .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1972, 222 (01) :59-+
[5]   FATAL PULMONARY HEMORRHAGE COMPLICATING USE OF A FLOW-DIRECTED BALLOON-TIPPED CATHETER IN A PATIENT RECEIVING ANTICOAGULANT THERAPY [J].
GOLDEN, MS ;
PINDER, T ;
ANDERSON, WT ;
CHEITLIN, MD .
AMERICAN JOURNAL OF CARDIOLOGY, 1973, 32 (06) :865-867
[6]   HEMOPTYSIS WITH FLOW-DIRECTED CARDIAC CATHETERIZATION [J].
LAPIN, ES ;
MURRAY, JA .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1972, 220 (09) :1246-&
[7]  
LEMEN R, 1975, NEW ENGL J MED, V292, P211
[8]   MULTIPURPOSE CATHETER FOR ELECTROPHYSIOLOGIC AND HEMODYNAMIC MONITORING PLUS ATRIAL-PACING [J].
MANTLE, JA ;
MASSING, GK ;
JAMES, TN ;
RUSSELL, RO ;
RACKLEY, CE .
CHEST, 1977, 72 (03) :285-290
[9]   PITFALLS OF SWAN-GANZ CATHETERIZATION [J].
SHIN, B ;
AYELLA, RJ ;
MCASLAN, TC .
CRITICAL CARE MEDICINE, 1977, 5 (03) :125-127
[10]   CATHETERIZATION OF HEART IN MAN WITH USE OF FLOW-DIRECTED BALLOON-TIPPED CATHETER [J].
SWAN, HJC ;
GANZ, W ;
FORRESTER, J ;
MARCUS, H ;
DIAMOND, G ;
CHONETTE, D .
NEW ENGLAND JOURNAL OF MEDICINE, 1970, 283 (09) :447-+