EVALUATION OF TRANS-THORACIC ELECTRICAL-IMPEDANCE IN THE DIAGNOSIS OF PULMONARY-EDEMA

被引:58
作者
FEIN, A
GROSSMAN, RF
JONES, JG
GOODMAN, PC
MURRAY, JF
机构
[1] SAN FRANCISCO GEN HOSP,MED CTR,MED SERV,SAN FRANCISCO,CA 94110
[2] SAN FRANCISCO GEN HOSP,MED CTR,RADIOL SERV,SAN FRANCISCO,CA 94110
[3] UNIV CALIF SAN FRANCISCO,CARDIOVASC RES INST,SAN FRANCISCO,CA 94143
[4] UNIV CALIF SAN FRANCISCO,DEPT MED,SAN FRANCISCO,CA 94143
[5] UNIV CALIF SAN FRANCISCO,DEPT RADIOL,SAN FRANCISCO,CA 94143
关键词
D O I
10.1161/01.CIR.60.5.1156
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To evaluate the clinical usefulness of measuring transthoracic electrical impedance in patients with pulmonary edema, we studied 27 normal subjects and 33 patients. In normal subjects, impedance increased when body position changed from supine to standing (p<0.01) and when lung volume increased from residual volume to total lung capacity (p<0.01). Impedance values varied (±2.7 Ω) when measured on 4 separate days but were relatively constant (±0.7 Ω) when measured hourly without removing the electrodes. In only five of 11 patients with clinically severe pulmonary edema and five of eight with radiologically severe pulmonary edema were impedance values outside the normal range. Changes in impedance values correlated with changes in clinical and radiographic indexes of pulmonary edema. We conclude that (1) single measurements of impedance are useless in diagnosing pulmonary edema, and (2) impedance changes as predicted as pulmonary edema increases and decreases, and these changes are readily detected by simpler methods of evaluation.