SIMULTANEOUS DETERMINATION OF ANIONS AND DIVALENT-CATIONS USING ION CHROMATOGRAPHY WITH ETHYLENEDIAMINETETRAACETIC ACID AS ELUENT

被引:21
作者
LEGRAS, CAA
机构
关键词
ION CHROMATOGRAPHY; ETHYLENEDIAMINETETRAACETIC ACID ELUENT; ULTRAVIOLET AND CONDUCTIVITY DETECTION; ANIONS AND DIVALENT CATIONS; FRESH-WATER AND SEA-WATER SAMPLES;
D O I
10.1039/an9931801035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method is described for the simultaneous determination of Cl-, NO3-, SO42-, Mg2+ and Ca2+ ions using ion chromatography with 0.0015 mol l-1 ethylenediaminetetraacetic acid as eluent at pH 4.8, with ultraviolet (UV) (200 nm) and conductivity detectors in series. Detection limits for these ions [2s (n = 15) for a 1 + 99 dilution of the standard solution] are 6,0.8,4,20 and 10 ng, respectively; a 200 mul aliquot was used for all determinations. In general, greater sensitivity is obtained for Cl-, SO42- and Mg2+ using conductivity detection, whereas for NO3- and Ca2+, UV detection affords a lower detection limit. The greater sensitivity for NO3- and Ca2+ compared with literature methods is due to a shorter wavelength for UV detection being accessible because of the lower absorbance of the eluent at pH 4.8. For Mg2+, partial protonation of the complex at this pH is thought to increase its equivalent conductance and hence sensitivity to conductivity detection. Analyses were made of four reference waters, which were within the 95% confidence interval for all analytes and for both detection modes. The calibration graphs were linear (r > 0.99) over at least three orders of magnitude of concentration from the detection limit, except for the UV detection of NO3- (r = 0.967). The method is directly applicable to both fresh- and sea-water samples, and was tested using a number of samples from natural and artificial water bodies, near Jabiru, Northern Territory, Australia. Reproducible analyses for all analytes were obtained for these freshwater samples, with alkalinities in the range 1-180 mg l-1 CaCO3, after storage at 4-degrees-C for 1 month. All peaks are well separated and no common ions interfere with any analyte.
引用
收藏
页码:1035 / 1041
页数:7
相关论文
共 27 条
[1]  
ANDEREGG G, 1977, IUPAC CHEM DATA SERI, V14, P3
[2]  
BARKLEY NP, 1978, ION CHROMATOGRAPHIC, V2, P115
[3]   AUTOMATED SIMULTANEOUS ANALYSIS OF ANIONS AND MONO-VALENT AND DIVALENT-CATIONS [J].
CHEAM, V ;
CHAU, ASY .
ANALYST, 1987, 112 (07) :993-997
[4]   ANION CHROMATOGRAPHY WITH LOW-CONDUCTIVITY ELUENTS [J].
GJERDE, DT ;
FRITZ, JS ;
SCHMUCKLER, G .
JOURNAL OF CHROMATOGRAPHY, 1979, 186 (DEC) :509-519
[5]   ANION CHROMATOGRAPHY WITH LOW-CONDUCTIVITY ELUENTS .2. [J].
GJERDE, DT ;
SCHMUCKLER, G ;
FRITZ, JS .
JOURNAL OF CHROMATOGRAPHY, 1980, 187 (01) :35-45
[6]   EFFECT OF (ETHYLENEDIAMINETETRAACETATO)COPPER(II) AND BIS(ETHYLENEDIAMINE)COPPER(II) ELUENTS ON NONSUPPRESSED ION CHROMATOGRAPHY WITH INDIRECT PHOTOMETRIC DETECTION [J].
HAYAKAWA, K ;
SAWADA, T ;
SHIMBO, K ;
MIYAZAKI, M .
ANALYTICAL CHEMISTRY, 1987, 59 (18) :2241-2245
[7]  
IKEDA S, 1991, NIPPON KAGAKU KAISHI, V6, P816
[8]   SIMULTANEOUS INDEPENDENT ANALYSIS OF ANIONS AND CATIONS USING INDIRECT PHOTOMETRIC CHROMATOGRAPHY [J].
ISKANDARANI, Z ;
MILLER, TE .
ANALYTICAL CHEMISTRY, 1985, 57 (08) :1591-1594
[9]   SINGLE INJECTION ION CHROMATOGRAPHIC ANALYSIS OF BOTH ANIONS AND CATIONS [J].
JONES, VK ;
TARTER, JG .
JOURNAL OF CHROMATOGRAPHY, 1984, 312 (NOV) :456-460
[10]   ION CHROMATOGRAPHIC DETERMINATION OF NITRATE AND SULFATE IN NATURAL-WATERS CONTAINING HUMIC SUBSTANCES [J].
MARKOVARGA, G ;
CSIKY, I ;
JONSSON, JA .
ANALYTICAL CHEMISTRY, 1984, 56 (12) :2066-2069