INCREASE IN DEFIBRILLATION THRESHOLD IN NON-THORACOTOMY IMPLANTABLE DEFIBRILLATORS USING A BIPHASIC WAVE-FORM

被引:36
作者
MARTIN, DT
JOHN, R
VENDITTI, FJ
机构
[1] Cardiac Electrophysiology Laboratory, Section of Cardiovascular Medicine, Lahey Clinic Medical Center, Burlington, MA
关键词
D O I
10.1016/S0002-9149(99)80078-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We have previously reported a chronic increase in defibrillation threshold in a non-thoracotomy implantable cardioverter-defibrillator (ICD) system using monophasic waveforms. To determine if this phenomenon is related to the lead system or the waveform used, we studied the chronic defibrillation threshold in consecutive patients receiving an ICD capable of delivering biphasic waveforms with the same lead system previously evaluated. Twenty-five patients received an ICD with biphasic shock waveform and have been followed for 4 to 15 months. All have undergone defibrillation threshold measurements using the identical testing protocol with biphasic waveforms at implant and at 2 months. Coronary artery disease was present in 15, idiopathic dilated cardiomyopathy in 9, and valvular heart disease in 1. The presenting arrhythmia was ventricular fibrillation in 11, ventricular tachycardia in 10, and syncope with inducible ventricular tachycardia in 4. The configuration of the shocking electrodes was randomized; the lead-only configuration was used in 14 patients (56%), and a subcutaneous patch was used in the remaining patients. Mean defibrillation threshold using a step-down technique was 9.8 +/- 1.0 J at implant, 13.2 +/- 1.6 J at 2 months, and 12.4 +/- 1.5 J at 6 months (p = 0.01 by analysis of variance). There was no change in clinical status, cardiac size, radiographic lead position, or impedance between implant and 2 months. These findings suggest the increase in defibrillation threshold in this ICD system is not related to the type of waveform used, but rather is a feature of non-thoracotomy as opposed to epicardial electrodes.
引用
收藏
页码:263 / 266
页数:4
相关论文
共 18 条
  • [1] RELATIONSHIP OF LEFT-VENTRICULAR MASS TO DEFIBRILLATION THRESHOLD FOR THE IMPLANTABLE DEFIBRILLATOR - A COMBINED CLINICAL AND ANIMAL STUDY
    CHAPMAN, PD
    SAGAR, KB
    WETHERBEE, JN
    TROUP, PJ
    [J]. AMERICAN HEART JOURNAL, 1987, 114 (02) : 274 - 278
  • [2] THE RELATIONSHIP BETWEEN SUCCESSFUL DEFIBRILLATION AND DELIVERED ENERGY IN OPEN-CHEST DOGS - REAPPRAISAL OF THE DEFIBRILLATION THRESHOLD CONCEPT
    DAVY, JM
    FAIN, ES
    DORIAN, P
    WINKLE, RA
    [J]. AMERICAN HEART JOURNAL, 1987, 113 (01) : 77 - 84
  • [3] EVALUATION OF ANTIARRHYTHMIC DRUGS ON DEFIBRILLATION ENERGY-REQUIREMENTS IN DOGS - SODIUM-CHANNEL BLOCK AND ACTION-POTENTIAL PROLONGATION
    ECHT, DS
    BLACK, JN
    BARBEY, JT
    COXE, DR
    CATO, E
    [J]. CIRCULATION, 1989, 79 (05) : 1106 - 1117
  • [4] LONG-TERM STABILITY OF DEFIBRILLATION THRESHOLDS WITH INTRAPERICARDIAL DEFIBRILLATOR PATCHES
    FRAME, R
    BRODMAN, R
    FURMAN, S
    GROSS, J
    KIM, SG
    FERRICK, K
    ROTH, J
    HOLLINGER, I
    FISHER, JD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01): : 208 - 212
  • [5] THE DEFIBRILLATION SUCCESS RATE VERSUS ENERGY RELATIONSHIP .1. CURVE FITTING AND THE MOST EFFICIENT DEFIBRILLATION ENERGY
    GLINER, BE
    MURAKAWA, Y
    THAKOR, NV
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (03): : 326 - 338
  • [6] EARLY CHANGES IN DEFIBRILLATION THRESHOLD FOLLOWING IMPLANTATION OF A NONTHORACOTOMY SYSTEM IN DOGS
    HALPERIN, BD
    HAUPT, DI
    LEMMER, JH
    HOLCOMB, SR
    OLIVER, RP
    SILKA, MJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 1771 - 1777
  • [7] EARLY POSTOPERATIVE INCREASE IN DEFIBRILLATION THRESHOLD WITH NONTHORACOTOMY SYSTEM IN HUMANS
    HSIA, HH
    MITRA, RL
    FLORES, BT
    MARCHLINSKI, FE
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (06): : 1166 - 1173
  • [8] HSIA HH, 1994, CIRCULATION, V90, P122
  • [9] CURRENT CONCEPTS FOR SELECTING THE LOCATION, SIZE AND SHAPE OF DEFIBRILLATION ELECTRODES
    IDEKER, RE
    WOLF, PD
    ALFERNESS, C
    KRASSOWSKA, W
    SMITH, WM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (02): : 227 - &
  • [10] DEFIBRILLATION EFFICACY - COMPARISON OF DEFIBRILLATION THRESHOLD VERSUS DOSE-RESPONSE CURVE DETERMINATION
    JONES, DL
    IRISH, WD
    KLEIN, GJ
    [J]. CIRCULATION RESEARCH, 1991, 69 (01) : 45 - 51