ULTRASOUND FOCAL LESION DETECTABILITY PHANTOMS

被引:30
作者
MADSEN, EL
ZAGZEBSKI, JA
MACDONALD, MC
FRANK, GR
机构
[1] UNIV WISCONSIN,DEPT HUMAN ONCOL,MADISON,WI 53706
[2] UNIV WISCONSIN,DEPT RADIOL,MADISON,WI 53706
关键词
ULTRASOUND; PHANTOM; DETECTABILITY; LESION; TISSUE-MIMICKING;
D O I
10.1118/1.596589
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two phantoms for assessing the performance of ultrasound scanners regarding detectability of small focal lesions are described. The spherical simulated lesions in the phantoms have diameters of 2.4, 3.0, and 4.0 mm and backscatter coefficients which are 16, 9, and 6 dB below that of the surrounding tissue-mimicking material, the latter simulating normal tissue such as liver. Random positioning and relatively large numbers of lesions deal with the statistical problem related to mistaking a normal textural fluctuation, characteristic of ultrasound images of parenchymal tissue, to be a small lesion. All lesions have ultrasonic properties that are the same as those of the surrounding material except for the backscatter coefficient; therefore, artifacts resulting from attenuation, reverberation, and/or refraction are absent, and detectability of deeper lesions is not influenced by the presence of more proximal ones. A method is outlined for using the phantoms to determine bounding proximal and distal depths of resolution, or detectability, of focal lesions of a given object contrast and size. Tests for observer dependence involving five subjects resulted in standard deviations in the distal and proximal depths of less than 0.5 cm. A repeat of the observer dependence tests 18 months later showed little change in the results. In addition, results obtained by two experienced observers are also given for a variety of specific scanners, scanning heads, and scanning parameters.
引用
收藏
页码:1171 / 1180
页数:10
相关论文
共 16 条
  • [1] QUANTITATIVE CONTRAST MEASUREMENTS IN B-MODE IMAGES COMPARISON BETWEEN EXPERIMENT AND THEORY
    BLY, SHP
    LEECHAHAL, D
    FOSTER, DR
    PATTERSON, MS
    FOSTER, FS
    HUNT, JW
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1986, 12 (03) : 197 - 208
  • [2] NEW ULTRASOUND TISSUE-EQUIVALENT MATERIAL
    BURLEW, MM
    MADSEN, EL
    ZAGZEBSKI, JA
    BANJAVIC, RA
    SUM, SW
    [J]. RADIOLOGY, 1980, 134 (02) : 517 - 520
  • [3] GOLDSTEIN A, 1988, J ULTRAS MED, V7, P487
  • [4] A CLINICAL-EVALUATION OF CONTRAST-DETAIL ANALYSIS FOR ULTRASOUND IMAGES
    LOPEZ, H
    LOEW, MH
    BUTLER, PF
    HILL, MC
    ALLMAN, RM
    [J]. MEDICAL PHYSICS, 1990, 17 (01) : 48 - 57
  • [5] Lopez H., 1988, Proceedings of the SPIE - The International Society for Optical Engineering, V914, P60, DOI 10.1117/12.968615
  • [6] METHOD OF DATA REDUCTION FOR ACCURATE DETERMINATION OF ACOUSTIC BACKSCATTER COEFFICIENTS
    MADSEN, EL
    INSANA, MF
    ZAGZEBSKI, JA
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1984, 76 (03) : 913 - 923
  • [7] OIL-IN-GELATIN DISPERSIONS FOR USE AS ULTRASONICALLY TISSUE-MIMICKING MATERIALS
    MADSEN, EL
    ZAGZEBSKI, JA
    FRANK, GR
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (03) : 277 - 287
  • [8] MADSEN EL, 1985, J ULTRAS MED, V4, pA47
  • [9] MORSE PM, 1953, METHODS THEORETICAL, P1066
  • [10] THE IMPROVEMENT AND QUANTITATIVE ASSESSMENT OF B-MODE IMAGES PRODUCED BY AN ANNULAR ARRAY CONE HYBRID
    PATTERSON, MS
    FOSTER, FS
    [J]. ULTRASONIC IMAGING, 1983, 5 (03) : 195 - 213