BLOOD-FLOW IMAGING THROUGH DETECTION OF TEMPORAL VARIATIONS IN MAGNETIZATION

被引:14
作者
WOLF, RL [1 ]
RICHARDSON, DB [1 ]
LAPLANTE, CC [1 ]
HUSTON, J [1 ]
RIEDERER, SJ [1 ]
EHMAN, RL [1 ]
机构
[1] MAYO CLIN & MAYO FDN, DEPT RADIOL, 200 1ST ST SW, ROCHESTER, MN 55905 USA
关键词
BLOOD; FLOW DYNAMICS; MAGNETIC RESONANCE (MR); IMAGE PROCESSING; VASCULAR STUDIES; PHANTOMS;
D O I
10.1148/radiology.185.2.1410373
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two hypotheses were tested: (a) that view-to-view variations in bulk phase and modulus of magnetization in vascular volume elements can indicate the presence of disordered blood flow, and (b) that a substantial loss of signal intensity on magnetic resonance (MR) angiograms of poststenotic regions is due to view-to-view changes in magnetization. To test these hypotheses, a technique was developed in which view-to-view variations in transverse magnetization were used to create angiographic projection images, which showed only disordered flow (disordered flow maps) in vitro and in vivo. In phantom studies, this technique recovered signal intensity downstream from stenoses. A combination of disordered flow maps with morphologic images improved visualization of stenotic regions and provided information on characteristics of local flow. These results show that view-to-view variations in transverse magnetization occur in regions of disordered flow and are an important cause of loss of signal intensity. This technique can provide information about dynamic blood flow and improve depiction of anatomic structures on MR angiograms.
引用
收藏
页码:559 / 567
页数:9
相关论文
共 37 条
[1]   BLOOD-FLOW EFFECTS IN MAGNETIC-RESONANCE IMAGING [J].
AXEL, L .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) :1157-1166
[2]  
AZUMA T, 1976, BIORHEOLOGY, V13, P337
[3]   THE APPEARANCE OF RAPIDLY FLOWING BLOOD ON MAGNETIC-RESONANCE IMAGES [J].
BRADLEY, WG ;
WALUCH, V ;
LAI, KS ;
FERNANDEZ, EJ ;
SPALTER, C .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) :1167-1174
[4]   MR ANGIOGRAPHY WITH OBLIQUE GRADIENT-RECALLED ECHO TECHNIQUE [J].
BROWN, DG ;
RIEDERER, SJ ;
JACK, CR ;
FARZANEH, F ;
EHMAN, RL .
RADIOLOGY, 1990, 176 (02) :461-466
[5]   EFFECTS OF DIFFUSION ON FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE EXPERIMENTS [J].
CARR, HY ;
PURCELL, EM .
PHYSICAL REVIEW, 1954, 94 (03) :630-638
[6]   DISORDER DISTAL TO MODELED STENOSES IN STEADY AND PULSATILE FLOW [J].
CASSANOVA, RA ;
GIDDENS, DP .
JOURNAL OF BIOMECHANICS, 1978, 11 (10-1) :441-453
[7]  
CHAO PW, 1989, BOOK ABSTRACTS, P235
[8]   CSF FLOW ARTIFACT REDUCTION USING CARDIAC CYCLE ORDERED PHASE-ENCODING METHOD [J].
CHO, MH ;
KIM, WS ;
CHO, ZH .
MAGNETIC RESONANCE IMAGING, 1990, 8 (04) :395-405
[9]   MR FOURIER-TRANSFORM ARTERIOGRAPHY USING SPECTRAL DECOMPOSITION [J].
CHO, ZH ;
JUNG, KJ ;
RO, YM .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) :226-237
[10]   MAGNETIC-RESONANCE ANGIOGRAPHY [J].
DUMOULIN, CL ;
HART, HR .
RADIOLOGY, 1986, 161 (03) :717-720