CHARACTERIZATION OF SPATIAL DISTORTION IN MAGNETIC-RESONANCE-IMAGING AND ITS IMPLICATIONS FOR STEREOTAXIC SURGERY

被引:202
作者
SUMANAWEERA, TS [1 ]
ADLER, JR [1 ]
NAPEL, S [1 ]
GLOVER, GH [1 ]
机构
[1] STANFORD UNIV,SCH MED,DEPT NEUROSURG,STANFORD,CA 94305
关键词
DISTORTION CORRECTION; MAGNETIC RESONANCE DISTORTION; SPATIAL ACCURACY; STEREOTAXIC SURGERY;
D O I
10.1227/00006123-199410000-00016
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
THE DIFFERENT SOURCES of spatial distortion in magnetic resonance images are reviewed from the point of view of stereotactic target localization. The extents of the two most complex sources of spatial distortion, gradient field nonlinearities and magnetic field inhomogeneities, are discussed both qualitatively and quantitatively. Several ways by which the spatial distortion resulting from these sources can be minimized are discussed. The clinical relevance of the spatial distortion along with some strategies to minimize the localization errors in magnetic resonance-guided stereotaxy are presented.
引用
收藏
页码:696 / 703
页数:8
相关论文
共 21 条
[1]   A TECHNIQUE FOR ACCURATE MAGNETIC-RESONANCE-IMAGING IN THE PRESENCE OF FIELD INHOMOGENEITIES [J].
CHANG, H ;
FITZPATRICK, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (03) :319-329
[2]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11
[3]  
CURRY T, 1990, CHRISTENSENS PHYSICS
[4]   SIMPLE PROTON SPECTROSCOPIC IMAGING [J].
DIXON, WT .
RADIOLOGY, 1984, 153 (01) :189-194
[5]  
DONG S, 1992, 5TH ANN IEEE S COMP
[6]  
GALLOWAY R L JR, 1991, Biomedical Instrumentation and Technology, V25, P457
[7]  
GAUVIN A, 1992, THESIS MCGILL U MONT
[8]  
Ludeke K M, 1985, Magn Reson Imaging, V3, P329, DOI 10.1016/0730-725X(85)90397-2
[9]  
MAURER CR, 1994, EFFECT GEOMETRIC DIS
[10]  
PYKETT IL, 1987, MAGNET RESON MED, V5, P563