DESIGN OF AN INTERVERTEBRAL-DISK PROSTHESIS

被引:81
作者
HEDMAN, TP
KOSTUIK, JP
FERNIE, GR
HELLIER, WG
机构
[1] Orthopaedic Biomechanics Laboratory, Toronto General Hospital, Toronto, ON
[2] Centre for Studies in Aging, Sunnybrook Medical Centre, Toronto, ON
关键词
INTERVERTEBRAL DISK PROSTHESIS; DESIGN CRITERIA; MOTION CONSTRAINTS;
D O I
10.1097/00007632-199106001-00016
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
This article presents criteria for the design of a lumbar intervertebral disc prosthesis that take into consideration issues of endurance, materials behavior, geometry, kinetics, motion constraints, fixation to bone, and safety. The criteria and design philosophies discussed are generally applicable in the synthesis of any new implant system. Specifications of the design of a disc prosthesis are presented along with an evaluation of the design based on some of the preliminary test data.
引用
收藏
页码:S256 / S260
页数:5
相关论文
共 21 条
  • [1] Andersson G., Epidemiologic aspects on low-back pain in industry, Spine, 6, pp. 53-60, (1981)
  • [2] Butlner-Janz K., Schellnack K., Zippel H., Modular type SB Charity lumbar intervertebral disk endoprosthesis-biomechanical and clinical results, Humboldt-Universitiit zu Berlin, GDR, Unpublished Data, (1989)
  • [3] Buttner-Janz K., Schellnack K., Zippel H., Biomechanics of the SB Charity lumbar intervertebral disc endoprosthesis, Intorthop, 13, pp. 173-176, (1989)
  • [4] Clark I.C., Site Visit-Toronto General Hospital, a Confidential Report on the Spinal Implant Project, (1987)
  • [5] Harris R.I., Wiley J.J., Acquired spondylolysis as a sequel to spine fusion, J Bone Joint Surg, 45A, (1963)
  • [6] Hcdman T.P., Kostuik J.P., Fernie G.R., Maki B., Artificial spinal disc. U.S. Patent, Filed, (1987)
  • [7] Hellier W.G., Spring-in-pocket simulation wear of an intervertebral disc prosthesis, Unpublished Data
  • [8] Hellier W.G., The effect of cobalt-chromium-molybdenum alloy powder plasma and sintered bead coatings on the fatigue response on hot isostatically pressed cobalt-chromium-molybdenum alloy powder substrates, Unpublished Data
  • [9] Hellier W.G., Hedman T.P., Pin-in-slot stimulation wear of an intervertebral disc prosthesis, Unpublished Data
  • [10] Hellier W.G., Hedman T.P., Wear volume rates of forged and hot isostatically pressed powder cobalt-chromium-molybdenum alloy for an intervertebral disc prosthesis, Unpublished Data