IN-VITRO AND IN-VIVO PERFORMANCE AND LIFETIME OF PERFLUORINATED IONOMER-COATED GLUCOSE SENSORS AFTER HIGH-TEMPERATURE CURING

被引:113
作者
MOUSSY, F
JAKEWAY, S
HARRISON, DJ
RAJOTTE, RV
机构
[1] UNIV ALBERTA, DEPT CHEM, EDMONTON T6G 2G2, AB, CANADA
[2] UNIV ALBERTA, SURG MED RES INST, EDMONTON T6G 2N8, AB, CANADA
关键词
D O I
10.1021/ac00094a007
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper we report that curing at 120 degrees C can be used to improve the in vivo durability of a miniaturized glucose sensor with an outer coating of the Dupont perfluorinated ionomer, Nafion. Sensors based on glucose oxidase trapped in an albumin/glutaraldehyde matrix were able to withstand curing at 120 degrees C without noticeable change in electrode sensitivity (+/-22% SD). Curing above 120 degrees C caused a gradual decline in sensitivity, with no sensitivity seen at 170 degrees C. Curing Nafion at 120 degrees C eliminated ascorbic acid and urea interferences and improved selectivity for glucose against uric acid and acetaminophen, compared to room temperature-cured Nafion coatings. The Nafion film reduced O-2 demand by the sensor, so the signal was O-2 independent across a partial pressure range of 8-140 mmHg. Several of the fully assembled, heat-cured, needle-type glucose sensors remained functional for at least 10 days after subcutaneous implantation in dogs, without degradation of their sensitivity (average 3 nA/mM in vivo at 37 degrees C and 6 nA/mM in vitro at 37 degrees C).
引用
收藏
页码:3882 / 3888
页数:7
相关论文
共 32 条
  • [1] CLINICAL CONTROL OF DIABETES BY ARTIFICIAL PANCREAS
    ALBISSER, AM
    LEIBEL, BS
    EWART, TG
    DAVIDOVAC, Z
    BOTZ, CK
    ZINGG, W
    SCHIPPER, H
    GANDER, R
    [J]. DIABETES, 1974, 23 (05) : 397 - 404
  • [2] DESIGN AND INVITRO STUDIES OF A NEEDLE-TYPE GLUCOSE SENSOR FOR SUBCUTANEOUS MONITORING
    BINDRA, DS
    ZHANG, YN
    WILSON, GS
    STERNBERG, R
    THEVENOT, DR
    MOATTI, D
    REACH, G
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (17) : 1692 - 1696
  • [3] BLUM L, 1991, BIOSENSORS PRINCIPLE
  • [4] FERROCENE-MEDIATED ENZYME ELECTRODE FOR AMPEROMETRIC DETERMINATION OF GLUCOSE
    CASS, AEG
    DAVIS, G
    FRANCIS, GD
    HILL, HAO
    ASTON, WJ
    HIGGINS, IJ
    PLOTKIN, EV
    SCOTT, LDL
    TURNER, APF
    [J]. ANALYTICAL CHEMISTRY, 1984, 56 (04) : 667 - 671
  • [5] NEEDLE ENZYME ELECTRODES FOR BIOLOGICAL STUDIES
    CHURCHOUSE, SJ
    BATTERSBY, CM
    MULLEN, WH
    VADGAMA, PM
    [J]. BIOSENSORS, 1986, 2 (06) : 325 - 342
  • [6] PERMEABILITY OF GLUCOSE AND OTHER NEUTRAL SPECIES THROUGH RECAST PERFLUOROSULFONATED IONOMER FILMS
    FAN, ZH
    HARRISON, DJ
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (11) : 1304 - 1311
  • [7] FISHER U, 1987, DIABETOLOGIA, V30, P940
  • [8] STRUCTURE AND RELATED PROPERTIES OF SOLUTION-CAST PERFLUOROSULFONATED IONOMER FILMS
    GEBEL, G
    ALDEBERT, P
    PINERI, M
    [J]. MACROMOLECULES, 1987, 20 (06) : 1425 - 1428
  • [9] CHARACTERIZATION OF PERFLUOROSULFONIC ACID POLYMER COATED ENZYME ELECTRODES AND A MINIATURIZED INTEGRATED POTENTIOSTAT FOR GLUCOSE ANALYSIS IN WHOLE-BLOOD
    HARRISON, DJ
    TURNER, RFB
    BALTES, HP
    [J]. ANALYTICAL CHEMISTRY, 1988, 60 (19) : 2002 - 2007
  • [10] GLUCOSE FAST-RESPONSE AMPEROMETRIC SENSOR BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN AN ELECTROPOLYMERIZED POLY(ORTHO-PHENYLENEDIAMINE) FILM
    MALITESTA, C
    PALMISANO, F
    TORSI, L
    ZAMBONIN, PG
    [J]. ANALYTICAL CHEMISTRY, 1990, 62 (24) : 2735 - 2740