NITROGEN-SPECIFIC LIQUID-CHROMATOGRAPHY DETECTOR BASED ON CHEMILUMINESCENCE - APPLICATION TO THE ANALYSIS OF AMMONIUM NITROGEN IN WASTE-WATER

被引:51
作者
FUJINARI, EM [1 ]
COURTHAUDON, LO [1 ]
机构
[1] ANTEK INSTRUMENTS GMBH,W-4000 DUSSELDORF 31,GERMANY
来源
JOURNAL OF CHROMATOGRAPHY | 1992年 / 592卷 / 1-2期
关键词
D O I
10.1016/0021-9673(92)85087-A
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel high-performance liquid chromatography-chemiluminescent nitrogen detection (HPLC-CLND) system is described. The analysis of ammonium nitrogen in metropolitan waste water by a new ion chromatographic method is presented. The significance of this novel nitrogen-specific detector is the capability for a routine and continuous operation. HPLC-CLND applications for anion and particularly reversed-phase techniques are also amenable. A nitrogen-specific detector for liquid chromatography opens a new dimension of analytical chemistry.
引用
收藏
页码:209 / 214
页数:6
相关论文
共 12 条
[1]   DETERMINATION OF AMMONIA AND METHYLAMINES IN AQUEOUS-SOLUTIONS BY ION CHROMATOGRAPHY [J].
BOUYOUCOS, SA .
ANALYTICAL CHEMISTRY, 1977, 49 (03) :401-403
[2]  
BRITTEN AJ, 1989, RES DEV, V31, P76
[3]  
COURTHAUDON LO, 1991, LC GC-MAG SEP SCI, V9, P732
[4]   HOMOGENEOUS CHEMILUMINESCENT MEASUREMENT OF NITRIC OXIDE WITH OZONE - IMPLICATIONS FOR CONTINUOUS SELECTIVE MONITORING OF GASEOUS AIR POLLUTANTS [J].
FONTIJN, A ;
SABADELL, AJ ;
RONCO, RJ .
ANALYTICAL CHEMISTRY, 1970, 42 (06) :575-&
[5]   CHEMI-LUMINESCENT NITROGEN DETECTOR GAS-CHROMATOGRAPHY AND ITS APPLICATION TO MEASUREMENT OF ATMOSPHERIC AMMONIA AND AMINES [J].
KASHIHIRA, N ;
MAKINO, K ;
KIRITA, K ;
WATANABE, Y .
JOURNAL OF CHROMATOGRAPHY, 1982, 239 (APR) :617-624
[6]   POLYMER-COATED CATION-EXCHANGE STATIONARY PHASES ON THE BASIS OF SILICA [J].
KOLLA, P ;
KOHLER, J ;
SCHOMBURG, G .
CHROMATOGRAPHIA, 1987, 23 (07) :465-472
[7]  
MILLER JM, 1988, CHROMATOGRAPHY CONCE, P182
[8]  
Parks R. E., 1977, U. S. Patent, Patent No. [4018562A, 4018562]
[9]   EVALUATION OF A NITROSYL-SPECIFIC GAS-PHASE CHEMI-LUMINESCENT DETECTOR WITH HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
ROBBAT, A ;
CORSO, NP ;
LIU, TY .
ANALYTICAL CHEMISTRY, 1988, 60 (02) :173-174
[10]   TRACE DETERMINATION OF AMINES AND OTHER NITROGEN CONTAINING COMPOUNDS WITH A MODIFIED THERMAL-ENERGY ANALYZER (TEA) [J].
ROUNBEHLER, DP ;
BRADLEY, SJ ;
CHALLIS, BC ;
FINE, DH ;
WALKER, EA .
CHROMATOGRAPHIA, 1982, 16 :354-358