KEYHOLE METHOD FOR ACCELERATING IMAGING OF CONTRAST AGENT UPTAKE

被引:476
作者
VANVAALS, JJ
BRUMMER, ME
DIXON, WT
TUITHOF, HH
ENGELS, H
NELSON, RC
GERETY, BM
CHEZMAR, JL
DENBOER, JA
机构
[1] Philips Medical Systems, Post-bus 10.000
[2] Philips Magnetic Resonance Research Center, Department of Radiology, Emory University School of Medicine, Atlanta
[3] Medisearch Foundation, En-Schede
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1993年 / 3卷 / 04期
关键词
CONTRAST ENHANCEMENT; FUNCTIONAL STUDIES; GADOLINIUM; IMAGE PROCESSING; KIDNEY; RAPID IMAGING;
D O I
10.1002/jmri.1880030419
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance (MR) imaging methods with good spatial and contrast resolution are often too slow to follow the uptake of contrast agents with the desired temporal resolution. Imaging can be accelerated by skipping the acquisition of data normally taken with strong phase-encoding gradients, restricting acquisition to weak-gradient data only. If the usual procedure of substituting zeros for the missing data is followed, blurring results. Substituting instead reference data taken before or well after contrast agent injection reduces this problem. Volunteer and patient images obtained by using such reference data show that imaging can be usefully accelerated severalfold. Cortical and medullary regions of interest and whole kidney regions were studied, and both gradient- and spin-echo images are shown. The method is believed to be compatible with other acceleration methods such as half-Fourier reconstruction and reading of more than one line of k space per excitation.
引用
收藏
页码:671 / 675
页数:5
相关论文
共 11 条
[1]  
Mansfield P, Pykett IL, Biological and medical imaging by NMR, J Magn Reson, 29, pp. 355-373, (1978)
[2]  
Belliveau JW, Kennedy DN, McKinstry RC, Et al., Functional mapping of the human visual cortex by magnetic resonance imaging, Science, 254, pp. 716-718, (1991)
[3]  
van Vaals JJ, Tuithof HH, Dixon WT, Increased time resolution in dynamic imaging (abstr), JMRI, 2, P, (1992)
[4]  
Liang ZP, Lauterbur PC, Improved temporal/spatial resolution in functional imaging through generalized series reconstruction (abstr), Works in progress supplement to annual meeting program, (1992)
[5]  
Bishop JE, Soutar I, Kucharczyk W, Plewes DB, Rapid sequential imaging with shared‐echo fast spin‐echo MR imaging (abstr), Works in progress supplement to annual meeting program, (1992)
[6]  
van Vaals JJ, Engels H, de Graaf RG, Et al., Method for accelerated perfusion imaging (abstr), Book of abstracts: Society of Magnetic Resonance in Medicine 1992, (1992)
[7]  
Brummer ME, Dixon WT, Gerety B, Tuithof HH, Composite k‐space windows (keyhole techniques) to improve temporal resolution in a dynamic series of images following contrast administration (abstr), Book of abstracts: Society of Magnetic Resonance in Medicine 1992, (1992)
[8]  
Liang ZP, Lauterbur PC, Efficient time‐sequential imaging through generalized series modeling: a simulation analysis (abstr), Book of abstracts: Society of Magnetic Resonance in Medicine 1992, (1992)
[9]  
Jones RA, Haraldseth O, Miller TB, Rinck PA, Unsgard G, Oksendal AN, Dynamic contrast enhanced, NMR perfusion imaging of regional cerebral ischemia in rats using k space substitution (abstr), Book of abstracts: Society of Magnetic Resonance in Medicine 1992, (1992)
[10]  
Tretter SA, Introduction to discrete time signal processing, pp. 227-233, (1976)