IMPROVED ACCURACY OF CONTINUOUS MEASUREMENT OF ARTERIAL OXYGEN-TENSION IN SICK NEWBORN-INFANTS - CRITERIA FOR READING AND RECALIBRATING THE ELECTRODE

被引:7
作者
WILKINSON, AR
PHIBBS, RH
GREGORY, GA
机构
[1] UNIV CALIF SAN FRANCISCO,DEPT PEDIAT,SAN FRANCISCO,CA 94143
[2] UNIV CALIF SAN FRANCISCO,DEPT ANESTHESIA,SAN FRANCISCO,CA 94143
[3] UNIV CALIF SAN FRANCISCO,CARDIOVASC RES INST,SAN FRANCISCO,CA 94143
关键词
D O I
10.1136/adc.54.4.307
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Improved accuracy of continuous arterial oxygen tension (PaO2) measurement is obtained if allowance is made for the response time of the electrode. Recalibration is necessary only when PaO2 measured in a blood sample differs from the electrode value by more than ±11.5 mmHg (1.5 kPa).
引用
收藏
页码:307 / 310
页数:4
相关论文
共 6 条
[1]   USE OF A CONTINUOUSLY RECORDING INTRAVASCULAR OXYGEN ELECTRODE IN NEWBORN [J].
GODDARD, P ;
KEITH, I ;
MARCOVITCH, H ;
ROBERTON, NR ;
ROLFE, P ;
SCOPES, JW .
ARCHIVES OF DISEASE IN CHILDHOOD, 1974, 49 (11) :853-860
[2]   CONTINUOUS ARTERIAL OXYGEN-TENSION MONITORING IN NEWBORN INFANT [J].
HARRIS, TR ;
NUGENT, M .
JOURNAL OF PEDIATRICS, 1973, 82 (06) :929-939
[3]  
KITTERMAN JA, 1970, PEDIATR CLIN N AM, V17, P895
[4]   COMPLICATIONS OF UMBILICAL OXYGEN ELECTRODES [J].
KOLLMEYER, KR ;
TSANG, RC .
JOURNAL OF PEDIATRICS, 1974, 84 (06) :894-897
[5]  
PARKER D, 1971, LANCET, V1, P952
[6]  
ROLFE P, 1976, MEDICAL ELECTRONICS