EXPERIMENTAL-DETERMINATION OF WALL CORRECTION FACTORS .2. NACP AND MARKUS PLANE-PARALLEL IONIZATION CHAMBERS

被引:16
作者
WITTKAMPER, FW [1 ]
AALBERS, AHL [1 ]
MIJNHEER, BJ [1 ]
机构
[1] NETHERLANDS MEASUREMENTS INST,3720 BA BILTHOVEN,NETHERLANDS
关键词
D O I
10.1088/0031-9155/37/4/013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pt.I see ibid., vol.35, p.835-46 (1990). The formalism to derive the absorbed dose to water from ionization chamber measurements in a phantom includes several wall correction factors, katt, km and pwall, that depend on the shape, size and composition of the ionization chamber. Wall correction factors have been measured for a number of NACP and PTW/Markus chambers. Significant deviations from calculated values occur due to uncertainties in the contribution to the total ionization from the different materials of these inhomogeneous ionization chambers. Experimental values of pwall larger than one have been obtained for the NACP chamber indicating that the side and back wall material of this chamber reduce the ionization compared with the situation where all electrons originate from the water phantom.
引用
收藏
页码:995 / 1004
页数:10
相关论文
共 20 条
[1]   CHAMBER-DEPENDENT WALL CORRECTION FACTORS IN DOSIMETRY [J].
ANDREO, P ;
NAHUM, A ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (11) :1189-1199
[2]   STOPPING POWER DATA FOR HIGH-ENERGY PHOTON BEAMS [J].
ANDREO, P ;
BRAHME, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (08) :839-858
[3]  
[Anonymous], 1983, MED PHYS, V10, P741
[5]   NGAS DETERMINATION FOR A PARALLEL-PLATE ION-CHAMBER [J].
KRITHIVAS, G ;
RAO, SN .
MEDICAL PHYSICS, 1986, 13 (05) :674-677
[6]   DETERMINATIONS OF NGAS AND PREPL FACTORS FROM COMMERCIALLY AVAILABLE PARALLEL-PLATE CHAMBERS - AAPM TASK GROUP-21 PROTOCOL [J].
KUBO, H ;
KENT, LJ ;
KRITHIVAS, G .
MEDICAL PHYSICS, 1986, 13 (06) :908-912
[7]   CALIBRATION AND USE OF PLANE-PARALLEL IONIZATION CHAMBERS FOR THE DETERMINATION OF ABSORBED DOSE IN ELECTRON-BEAMS [J].
MATTSSON, LO ;
JOHANSSON, KA ;
SVENSSON, H .
ACTA RADIOLOGICA ONCOLOGY, 1981, 20 (06) :385-399
[8]  
MATTSSON LO, 1984, THESIS U GOTEBORG SW
[9]  
NAHUM AE, 1988, P IAEA S DOSIMETRY R, V1, P87
[10]   CALCULATED RESPONSE AND WALL CORRECTION FACTORS FOR IONIZATION CHAMBERS EXPOSED TO CO-60 GAMMA-RAYS [J].
NATH, R ;
SCHULZ, RJ .
MEDICAL PHYSICS, 1981, 8 (01) :85-93