THERMAL PROPERTIES OF CHOLESTEROL AND ESTIMATION OF IMPURITY BY DIFFERENTIAL SCANNING CALORIMETRY

被引:3
作者
GENT, WLG
机构
[1] Biochemistry Department, Guy's Hospital Medical School
来源
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS | 1969年 / 2卷 / 01期
关键词
D O I
10.1088/0022-3735/2/1/317
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Differential scanning calorimetry has been used to investigate the heat of fusion (ΔHf=7·12 kcal per g mol) and true melting temperature (T0=422·4°K) of pure cholesterol (99·5 mole%). In addition to analysis of the thermograph by an existing method a new and more rapid procedure has been used for evaluation of the purity of substances. Evidence is given that differential scanning calorimetry measurements at low scanning rates (0·5,1°min-1), followed by correction for premelting, are essential for estimation of ΔH f of substances not ultra-pure, at higher scanning rates large errors result. The entropy of fusion of cholesterol is similar to that of the simpler cholesteryl esters.
引用
收藏
页码:69 / &
相关论文
共 12 条
[1]   SPECIFIC HEATS OF NEMATIC SMECTIC AND CHOLESTERIC LIQUID CRYSTALS BY DIFFERENTIAL SCANNING CALORIMETRY [J].
BARRALL, EM ;
PORTER, RS ;
JOHNSON, JF .
JOURNAL OF PHYSICAL CHEMISTRY, 1967, 71 (04) :895-&
[2]   HEATS OF TRANSITION FOR SOME CHOLESTERYL ESTERS BY DIFFERENTIAL SCANNING CALORIMETRY [J].
BARRALL, EM ;
PORTER, RS ;
JOHNSON, JF .
JOURNAL OF PHYSICAL CHEMISTRY, 1967, 71 (05) :1224-&
[3]   THE CRYSTAL STRUCTURE OF CHOLESTERYL IODIDE [J].
CARLISLE, CH ;
CROWFOOT, D .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1945, 184 (996) :64-&
[4]  
DICKINSON HC, 1915, NAT BUR STAND B, V12, P49
[6]  
FURAWAKA GT, 1951, J RESEARCH NATIONA A, V47, P256
[7]  
GRAY AP, 1966, PERKIN ELMER INSTRUM, V16, P9
[8]  
GRAY AP, 1966, PERKIN ELMER INSTRUM, V16, P15
[9]  
OELSEN W, 1955, Z METALLKD, V46, P555
[10]   ANALYSIS OF TEMPERATURE-CONTROLLED SCANNING CALORIMETER [J].
ONEILL, MJ .
ANALYTICAL CHEMISTRY, 1964, 36 (07) :1238-&