BACTERIAL AND ENDOTOXIN PERMEABILITY OF HEMODIALYSIS MEMBRANES

被引:26
作者
BERNICK, JJ
PORT, FK
FAVERO, MS
BROWN, DG
机构
[1] UNIV MICHIGAN HOSP,DEPT INORGAN & PHYS CHEM,DIV NEPHROL,ANN ARBOR,MI 48109
[2] VET ADM HOSP,ANN ARBOR,MI 48105
[3] CTR DIS CONTROL,PHOENIX,AZ 85014
关键词
D O I
10.1038/ki.1979.154
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Dialysis fluids containing at least 107 bacteria per milliliter and as much as 12,500 ng of endotoxin equivalents per milliliter were dialyzed and ultrafiltered with three types of disposable hemodialyzers. Neither bacteria nor endotoxin, as measured by the Limulus lysate assay, was detected in the sterile compartment despite ultrafiltration. Under these favorable conditions for endotoxin transfer, the maximum transfer rate was calculated to be less than 3.5 ng of endotoxin equivalents per hour. At this rate, it is unlikely that pyrexia during hemodialysis is due to the transfer of endotoxin across an intact dialyzing membrane is maintained, this investigation indicates that the risk of endotoxemia or bacteremia associated with the use of contaminated dialysis fluids is negligible.
引用
收藏
页码:491 / 496
页数:6
相关论文
共 35 条