EVALUATION OF A NEW VUV SOFT-X-RAY TOROIDAL GRATING MONOCHROMATOR WITH A MOVABLE EXIT SLIT

被引:11
作者
MANCINI, DC
BISSEN, M
RIOUX, D
PATEL, R
ROGERS, G
BRODSKY, EL
HOCHST, H
机构
[1] Synchrotron Radiation Center, University of Wisconsin Madison, Stoughton
关键词
D O I
10.1063/1.1143097
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The performance of a toroidal grating monochromator can be significantly improved by compensating for the wavelength dependent defocusing aberration by means of a movable exit slit. Model calculations of a movable exist slit design show a resolving power that varies nearly linearly with wavelength over the scan range. The energy resolution follows the slope between two optimized resolution minima of a fixed slit TGM. The overall improvement in resolution due to the movable slit depends critically on the figure error of the toroid surface. The Al(L(II,III)) core edges provide a means to calibrate gratings and determine the resolution at 72.8 eV. A method is described for using the shift in core edge with exit slit position to enable final alignment of the exit slit motion providing stability of energy calibration of 2 meV over the scan range. Comparison of calculations and measured resolution suggests that slope errors remain the limiting factor in performance.
引用
收藏
页码:1269 / 1273
页数:5
相关论文
共 16 条
[1]   THE THEORY OF THE CONCAVE GRATING [J].
BEUTLER, HG .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1945, 35 (05) :311-350
[2]   METAL PHOTOCATHODES AS SECONDARY STANDARDS FOR ABSOLUTE INTENSITY MEASUREMENTS IN VACUUM ULTRAVIOLET [J].
CAIRNS, RB ;
SAMSON, JAR .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1966, 56 (11) :1568-&
[3]   THE STUDY AND DESIGN OF A HIGH TRANSMISSION, HIGH-RESOLUTION TOROIDAL GRATING MONOCHROMATOR FOR SOFT-X-RAY RADIATION [J].
CHEN, CT ;
PLUMMER, EW ;
HOWELLS, MR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 222 (1-2) :103-106
[4]   6-M TGM IMPLEMENTATION AT THE WISCONSIN-SYNCHROTRON-RADIATION-CENTER (SRC) [J].
COLE, RK ;
PERKINS, FK ;
BRODSKY, EL ;
FILIPPONI, A ;
KORPELLA, E ;
MANCINI, DC ;
PRUETT, CH ;
WALLACE, DJ ;
WELNAK, JT ;
ZANINI, F .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :2093-2096
[5]   THE TORUS GRATING [J].
HABER, H .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1950, 40 (03) :153-165
[6]  
HENKE BL, 1981, J APPL PHYS, V52, P302
[7]   1ST RESULTS FROM A 6 M/10 M TOROIDAL GRATING MONOCHROMATOR FOR SOFT X-RAYS [J].
HIMPSEL, FJ ;
JUGNET, Y ;
EASTMAN, DE ;
DONELON, JJ ;
GRIMM, D ;
LANDGREN, G ;
MARX, A ;
MORAR, JF ;
ODEN, C ;
POLLAK, RA ;
SCHNEIR, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 222 (1-2) :107-110
[8]  
KAO CC, 1989, BNL42689 REP
[9]  
LARSEN PK, 1982, NUCL INSTRUM METHODS, V195, P233
[10]  
LENTH W, 1975, DESY F417507 INT BER