LARGE AND SMALL COILS IN MAGNETIC STIMULATION OF THE PROXIMAL PORTION OF THE FACIAL-NERVE

被引:3
作者
OISHI, M [1 ]
TAKASU, T [1 ]
机构
[1] NIHON UNIV,SCH MED,DEPT NEUROL,TOKYO 101,JAPAN
关键词
MAGNETIC STIMULATION; FACIAL NERVE; MAGNETIC COIL SIZE;
D O I
10.1159/000116897
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Magnetic stimulation (Magstim model 200) was performed with a large coll (14 cm in diameter) and a small coil (7 cm in diameter) in 12 healthy persons. Muscle potentials were obtained from ipsilateral orbicularis oris and nasalis muscles when magnetic stimulation was performed by placing a large or small magnetic coil at T6 or by placing a large magnetic coll at C4. The latency obtained by placing a magnetic coll at T6 was 1.2-1.3 ms (with a large coll) and 1.3-1.4 ms (with a small coil) longer, and the latency obtained by placing a large magnetic coil at C4 was 1.3-1.4 ms longer than the latencies obtained by stimulating at the stylomastoid foramen. The proximal portion of the facial nerve is considered to be stimulated transcranially when magnetic stimulation is performed by placing either a large or small coil at T6 or by placing a large coil at C4.
引用
收藏
页码:34 / 37
页数:4
相关论文
共 26 条
[1]  
BARKER AT, 1990, MAGNETIC STIMULATION, P55
[2]   TRANSCRANIAL MAGNETIC STIMULATION OF THE HUMAN-BRAIN - RESPONSES IN MUSCLES SUPPLIED BY CRANIAL NERVES [J].
BENECKE, R ;
MEYER, BU ;
SCHONLE, P ;
CONRAD, B .
EXPERIMENTAL BRAIN RESEARCH, 1988, 71 (03) :623-632
[3]  
BENECKE R, 1988, EEG-EMG-Z ELEK ELEKT, V19, P228
[4]  
CHOKROVERTY S, 1989, Electromyography and Clinical Neurophysiology, V29, P409
[5]  
CHOKROVERTY S, 1990, MAGNETIC STIMULATION, P249
[6]  
Cocito D, 1990, Acta Otorhinolaryngol Ital, V10, P475
[7]   HEARING-LOSS FROM THE ACOUSTIC ARTIFACT OF THE COIL USED IN EXTRACRANIAL MAGNETIC STIMULATION [J].
COUNTER, SA ;
BORG, E ;
LOFQVIST, L ;
BRISMAR, T .
NEUROLOGY, 1990, 40 (08) :1159-1162
[8]   MAGNETIC TRANSCRANIAL AND ELECTRIC DIRECT STIMULATION OF THE FACIAL MOTOR CORTEX IN DOGS [J].
ESTREM, S ;
HAGHIGHI, S ;
DAVIS, WE .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1990, 103 (03) :439-442
[9]   A COMPARISON OF MAGNETIC AND ELECTRICAL-STIMULATION OF FACIAL-NERVE AT THE CEREBELLOPONTINE ANGLE IN THE DOG [J].
ESTREM, SA ;
MCCORMACK, T ;
HAGHIGHI, SS ;
POTTER, T .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 75 (06) :558-560
[10]   THE UTILITY OF MAGNETIC STIMULATION FOR ROUTINE PERIPHERAL-NERVE CONDUCTION STUDIES [J].
EVANS, BA ;
LITCHY, WJ ;
DAUBE, JR .
MUSCLE & NERVE, 1988, 11 (10) :1074-1078