COMPUTER-CONTROLLED MULTIELEMENT ATOMIC EMISSION-FLUORESCENCE SPECTROMETER SYSTEM

被引:22
作者
ULLMAN, AH [1 ]
POLLARD, BD [1 ]
BOUTILIER, GD [1 ]
BATEH, RP [1 ]
HANLEY, P [1 ]
WINEFORDNER, JD [1 ]
机构
[1] UNIV FLORIDA,DEPT CHEM,GAINESVILLE,FL 32611
关键词
D O I
10.1021/ac50050a021
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A versatile system for multielement analysis and analytical studies of noise sources, flames, modulation techniques, statistics of error propagation has been constructed. The instrument consisted of an EIMAC xenon arc lamp for fluorescence excitation, a flame (most commonly air/acetylene or nitrous oxide/acetylene), a wavelength modulated, slewed-scan monochromator with data acquisition by a photon counter. The entire instrument was under computer (PDP 11) control. Analytical figures of merit (detection limits and growth curves) for 25 elements are presented along with a discussion of instrumental design, flexibility, mode of modulation, operation. Applicability to real samples Is demonstrated with respect to trace metal analysis in jet engine oils (SOAP samples) and NBS Standard Reference Materials. © 1979, American Chemical Society. All rights reserved.
引用
收藏
页码:2382 / 2387
页数:6
相关论文
共 19 条
[1]   REVIEW AND TUTORIAL DISCUSSION OF NOISE AND SIGNAL-TO-NOISE RATIOS IN ANALYTICAL SPECTROMETRY .2. FUNDAMENTAL PRINCIPLES OF SIGNAL-TO-NOISE RATIOS [J].
BOUTILIER, GD ;
POLLARD, BD ;
WINEFORDNER, JD ;
CHESTER, TL ;
OMENETTO, N .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1978, 33 (08) :401-416
[2]  
DEAN JA, 1971, FLAME EMISSION ATOMI, V2
[3]   WAVELENGTH MODULATION FOR BACKGROUND CORRECTION IN GRAPHITE FURNACE ATOMIC EMISSION SPECTROMETRY [J].
EPSTEIN, MS ;
RAINS, TC ;
OHAVER, TC .
APPLIED SPECTROSCOPY, 1976, 30 (03) :324-329
[4]   DOUBLE MODULATION ATOMIC FLUORESCENCE FLAME SPECTROMETRY [J].
FOWLER, WK ;
KNAPP, DO ;
WINEFORD.JD .
ANALYTICAL CHEMISTRY, 1974, 46 (04) :601-602
[5]  
FUJIWARA K, UNPUBLISHED
[6]  
HIEFTJE GM, 1978, ANAL CHEM, V50, pR300, DOI 10.1021/ac50028a029
[7]   MULTIELEMENT ANALYSIS VIA COMPUTER-CONTROLLED RAPID-SCAN ATOMIC FLUORESCENCE SPECTROMETER WITH A CONTINUUM SOURCE [J].
JOHNSON, DJ ;
PLANKEY, FW ;
WINEFORDNER, JD .
ANALYTICAL CHEMISTRY, 1975, 47 (11) :1739-1743
[8]  
JOHNSON DJ, 1974, ANAL CHEM, V46, P1898
[9]   EFFECT OF MODULATION WAVE FORM ON UTILITY OF EMISSION BACKGROUND CORRECTIONS OBTAINED WITH AN OSCILLATING REFRACTOR PLATE [J].
KOIRTYOHANN, SR ;
GLASS, ED ;
YATES, DA ;
HINDERBERGER, EJ ;
LICHTE, FE .
ANALYTICAL CHEMISTRY, 1977, 49 (08) :1121-1126
[10]  
MAVRODINEANU R, 1977, NBS492 SPEC PUBL