DETERMINATION OF LEAD IN BIOLOGICAL-MATERIALS BY MICROWAVE-ASSISTED MINERALIZATION AND FLOW-INJECTION ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY

被引:48
作者
BURGUERA, JL
BURGUERA, M
机构
[1] Department of Chemistry, Faculty of Sciences, University of Los Andes, Mérida 5101-A
关键词
LEAD DETERMINATION; BIOLOGICAL SAMPLE; FLOW INJECTION; ELECTROTHERMAL ATOMIC ABSORPTION SPECTROMETRY; ONLINE MINERALIZATION;
D O I
10.1039/ja9930800235
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The design and operation of an on-line automated microwave-assisted mineralization and flow injection electrothermal atomic absorption spectrometric system for the determination of lead in biological materials are described. Certified reference materials were homogeneously dispersed in a 0.4% v/v Triton X-100 solution with an ultrasonic device. Mineralization of samples was accomplished with an HCl-HNO3 plug downstream in a PTFE coil located inside the microwave oven. A gas diffusion cell device allowed efficient degasification of mineralized samples. Precise volumes of the mineralized samples, collected in a capillary of a sampling arm assembly, were introduced by means of positive displacement with air through a time-based solenoid injector into the graphite furnace. The entire system was controlled by a computer, independent of the spectrometer. The spectrometer autosampler used for the introduction of the chemical modifier was pre-programmed to synchronize with the operation of the flow system. The linear range was from 0 to 70 mug l-1 of lead and the detection limit was 0.1 mug l-1 (0.8 pg). The agreement between observed and certified values was good, except for the botanical samples Olive Leaves and Pine Needles (with an average recovery of about 900%). The main advantage of the method is that in this totally closed system with on-line sample mineralization low and repeatable blanks were obtained, allowing the rapid determination (about ten samples per hour) of lead in various biological matrices with minimum exposure to the environment and minimum operator attention.
引用
收藏
页码:235 / 241
页数:7
相关论文
共 34 条
  • [1] ALVARADO J, 1991, J TRACE ELEM ELECT H, V5, P173
  • [2] DESIGN OF A CONTINUOUS-FLOW 2-STEP EXTRACTION SAMPLE WORK-UP SYSTEM FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY
    BACKSTROM, K
    DANIELSSON, LG
    [J]. ANALYTICA CHIMICA ACTA, 1990, 232 (02) : 301 - 315
  • [3] BORJA RM, 1989, QUIM ANAL, V8, P241
  • [4] BROWN AA, 1987, J ANAL ATOM SPECTROM, V3, pR43
  • [5] Burguera J L, 1987, J Trace Elem Electrolytes Health Dis, V1, P21
  • [6] BURGUERA JL, 1992, ATOM SPECTROSC, V13, P67
  • [7] SIMPLE VARIABLE-VOLUME INJECTOR FOR FLOW-INJECTION SYSTEMS
    BURGUERA, JL
    BURGUERA, M
    RIVAS, C
    DELAGUARDIA, M
    SALVADOR, A
    [J]. ANALYTICA CHIMICA ACTA, 1990, 234 (01) : 253 - 257
  • [8] BURGUERA JL, 1990, J FLOW INJECTION ANA, V7, P11
  • [9] DETERMINATION OF ZINC AND CADMIUM IN SMALL AMOUNTS OF BIOLOGICAL TISSUES BY MICROWAVE-ASSISTED DIGESTION AND FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY
    BURGUERA, M
    BURGUERA, JL
    ALARCON, OM
    [J]. ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) : 421 - 427
  • [10] FLOW-INJECTION AND MICROWAVE-OVEN SAMPLE DECOMPOSITION FOR DETERMINATION OF COPPER, ZINC AND IRON IN WHOLE-BLOOD BY ATOMIC-ABSORPTION SPECTROMETRY
    BURGUERA, M
    BURGUERA, JL
    ALARCON, OM
    [J]. ANALYTICA CHIMICA ACTA, 1986, 179 : 351 - 357