IMPROVEMENT OF THE SPHERICAL MIRROR ANALYZER

被引:64
作者
DAIMON, H
INO, S
机构
[1] Department of Physics, Faculty of Science, University of Tokyo, Bunkyo-ku
关键词
D O I
10.1063/1.1141923
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
With the use of some obstacle rings and a set of retarding grids, the resolution of the prototype spherical mirror analyzer, which has been developed in a previous paper [H. Daimon, Rev. Sci. Instrum. 59, 545 (1988)], was improved. The uniformity of the resolution with respect to the emission angle was also achieved. Some examples of electron energy-loss spectra and XPS spectrum were presented.
引用
收藏
页码:57 / 60
页数:4
相关论文
共 6 条
[1]  
Daimon H., 1988, Journal of the Vacuum Society of Japan, V31, P954, DOI 10.3131/jvsj.31.954
[3]   AN ELLIPSOIDAL MIRROR DISPLAY ANALYZER SYSTEM FOR ELECTRON-ENERGY AND ANGULAR MEASUREMENTS [J].
EASTMAN, DE ;
DONELON, JJ ;
HIEN, NC ;
HIMPSEL, FJ .
NUCLEAR INSTRUMENTS & METHODS, 1980, 172 (1-2) :327-336
[4]   NONDISPERSIVE ELECTRON-ENERGY ANALYZER FOR ESCA [J].
LEE, JD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1973, 44 (07) :893-898
[5]   SOFT-X-RAY IMAGING ZONE PLATES WITH LARGE ZONE NUMBERS FOR MICROSCOPIC AND SPECTROSCOPIC APPLICATIONS [J].
NIEMANN, B ;
RUDOLPH, D ;
SCHMAHL, G .
OPTICS COMMUNICATIONS, 1974, 12 (02) :160-163
[6]   MORE ON SPHERICAL CONDENSER AS AN ANALYSER .I. NONRELATIVISTIC PART [J].
SAREL, HZ .
NUCLEAR INSTRUMENTS & METHODS, 1966, 42 (01) :71-&