HELICAL CT - PRINCIPLES AND TECHNICAL CONSIDERATIONS

被引:84
作者
BRINK, JA
HEIKEN, JP
WANG, G
MCENERY, KW
SCHLUETER, FJ
VANNIER, MW
机构
[1] Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, 63110., MO
关键词
COMPUTED TOMOGRAPHY (CT); HELICAL TECHNOLOGY; PHYSICS;
D O I
10.1148/radiographics.14.4.7938775
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The performance of helical CT requires several user-defined parameters that exceed the requirements of conventional CT. One needs to carefully select the collimation, table increment, and reconstruction interval. Minimizing these parameters maximizes longitudinal resolution but with various trade-offs. Decreasing the collimation decreases the effective section thickness but increases pixel noise. Limiting the table increment to a pitch of 1 limits the broadening of the effective section thickness associated with the helical technique but also limits the coverage that can be achieved with a given helical scan. Our general practice is to minimize the collimation to cover the volume of interest and to accept a pitch up to 2, provided that we are using 180 degree interpolation. The reconstruction interval is also minimized to maximize longitudinal resolution but with trade-offs of increased image processing time, data storage requirements, and physician time for image review. For routine diagnosis, we recommend reconstruction of one to two sections per table increment, and, for multiplanar and three-dimensional imaging, we recommend at least three sections be reconstructed per table increment. The scan duration is dictated by both patient and machine factors.
引用
收藏
页码:887 / 893
页数:7
相关论文
共 15 条
  • [1] Bresler Y., 1989, P ICASSP, V3, P1472, DOI [10.1109/ICASSP.1989.266718, DOI 10.1109/ICASSP.1989.266718]
  • [2] SPIRAL CT - DECREASED SPATIAL-RESOLUTION INVIVO DUE TO BROADENING OF SECTION-SENSITIVITY PROFILE
    BRINK, JA
    HEIKEN, JP
    BALFE, DM
    SAGEL, SS
    DICROCE, J
    VANNIER, MW
    [J]. RADIOLOGY, 1992, 185 (02) : 469 - 474
  • [3] PULMONARY NODULE - EVALUATION WITH SPIRAL VOLUMETRIC CT
    COSTELLO, P
    ANDERSON, W
    BLUME, D
    [J]. RADIOLOGY, 1991, 179 (03) : 875 - 876
  • [4] COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION
    CRAWFORD, CR
    KING, KF
    [J]. MEDICAL PHYSICS, 1990, 17 (06) : 967 - 982
  • [5] SPIRAL CT OF THE PANCREAS WITH MULTIPLANAR DISPLAY
    FISHMAN, EK
    WYATT, SH
    NEY, DR
    KUHLMAN, JE
    SIEGELMAN, SS
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 159 (06) : 1209 - 1215
  • [6] SPIRAL(HELICAL) CT
    HEIKEN, JP
    BRINK, JA
    VANNIER, MW
    [J]. RADIOLOGY, 1993, 189 (03) : 647 - 656
  • [7] SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION
    KALENDER, WA
    SEISSLER, W
    KLOTZ, E
    VOCK, P
    [J]. RADIOLOGY, 1990, 176 (01) : 181 - 183
  • [8] MCENERY KW, IN PRESS INVEST RADI
  • [9] MEASUREMENT OF SLICE SENSITIVITY PROFILES IN SPIRAL CT
    POLACIN, A
    KALENDER, WA
    BRINK, J
    VANNIER, MA
    [J]. MEDICAL PHYSICS, 1994, 21 (01) : 133 - 140
  • [10] EVALUATION OF SECTION SENSITIVITY PROFILES AND IMAGE NOISE IN SPIRAL CT
    POLACIN, A
    KALENDER, WA
    MARCHAL, G
    [J]. RADIOLOGY, 1992, 185 (01) : 29 - 35