Correction of eddy current-induced artefacts in diffusion tensor imaging using iterative cross-correlation

被引:80
作者
Bastin, ME [1 ]
机构
[1] Univ Edinburgh, Western Gen Hosp, Dept Med Phys & Med Engn, Edinburgh EH4 2XU, Midlothian, Scotland
关键词
diffusion tensor imaging; echo-planar imaging; eddy currents; Monte Carlo simulations;
D O I
10.1016/S0730-725X(99)00026-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Geometric distortions of echo-planar images produced by the strong eddy currents present in the diffusion tensor imaging experiment are a major confound to the accurate quantification of diffusion coefficients, and measures of diffusion anisotropy based upon them. Here rye investigate how the method of iterative cross-correlation (ICC) of baseline and diffusion-weighted images (DWIs) originally proposed by Haselgrove and Moore (Magn. Reson, Med. 36:960-964; 1996) can be extended to correct high b-value DWIs, without the need for extrapolation of distortion parameters determined from low b-value images. Monte Carlo simulations of synthetic brain images show that the maximum value of the trace of the b-matrix, Tr(b), at which distorted DWIs can be accurately corrected by direct comparison with the undistorted baseline image is approximately 300 s mm(-2). Removal of the cerebrospinal fluid signal greatly extends this value of Tr(b) (up to approximately 2000 s mm(-2)), thereby allowing direct comparison of baseline and distorted images. The use of ICC distortion parameters determined from separate calibrations of water phantom images is also investigated, and found to be effective in correcting geometric distortions observed in the DWIs collected as part of a human volunteer diffusion tensor imaging study. This work suggests that distorted DWIs acquired at high values of b may be corrected using the ICC algorithm without collecting additional low b-value images, thus allowing simplified methods of measuring the apparent diffusion tensor D, based on collecting a small number of DWIs, to be implemented in quantitative patient examinations. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:1011 / 1024
页数:14
相关论文
共 30 条
  • [1] Elimination of eddy current artifacts in diffusion-weighted echo-planar images: The use of bipolar gradients
    Alexander, AL
    Tsuruda, JS
    Parker, DL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) : 1016 - 1021
  • [2] Diffusion anisotropy measurements in ischaemic stroke of the human brain
    Armitage, P. A.
    Bastin, M. E.
    Marshall, I.
    Wardlaw, J. M.
    Cannon, J.
    [J]. MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 1998, 6 (01) : 28 - 36
  • [3] A simplified method to measure the diffusion tensor from seven MR images
    Basser, PJ
    Pierpaoli, C
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (06) : 928 - 934
  • [4] Inferring microstructural features and the physiological state of tissues from diffusion-weighted images
    Basser, PJ
    [J]. NMR IN BIOMEDICINE, 1995, 8 (7-8) : 333 - 344
  • [5] Basser PJ, 1996, J MAGN RESON SER B, V111, P209, DOI [10.1006/jmrb.1996.0086, 10.1016/j.jmr.2011.09.022]
  • [6] ESTIMATION OF THE EFFECTIVE SELF-DIFFUSION TENSOR FROM THE NMR SPIN-ECHO
    BASSER, PJ
    MATTIELLO, J
    LEBIHAN, D
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES B, 1994, 103 (03): : 247 - 254
  • [7] A theoretical study of the effect of experimental noise on the measurement of anisotropy in diffusion imaging
    Bastin, ME
    Armitage, PA
    Marshall, I
    [J]. MAGNETIC RESONANCE IMAGING, 1998, 16 (07) : 773 - 785
  • [8] The use of diffusion tensor imaging in quantifying the effect of dexamethasone on brain tumours
    Bastin, ME
    Delgado, M
    Whittle, IR
    Cannon, J
    Wardlaw, JM
    [J]. NEUROREPORT, 1999, 10 (07) : 1385 - 1391
  • [9] Calamante F, 1999, MAGN RESON MED, V41, P95, DOI 10.1002/(SICI)1522-2594(199901)41:1<95::AID-MRM14>3.0.CO
  • [10] 2-T