ACCEPTANCE TESTING PRONE STEREOTAXIC BREAST BIOPSY UNITS

被引:8
作者
KIMMESMITH, C
SOLBERG, T
机构
关键词
MAMMOGRAPHY INSTRUMENTATION; MAMMOGRAPHY QUALITY CONTROL; BREAST BIOPSY EQUIPMENT;
D O I
10.1118/1.597353
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
When the Mammography Quality Standards Act becomes law in October, 1994, stereotactic breast biopsy units may require yearly physicist calibration. Upright stereotactic units can be easily tested using conventional mammography procedures and a gelatin phantom containing simulated calcifications, but prone units are difficult to assess because of the under-table tube configuration. The two current manufacturers of these units have made different design decisions which affect each unit's calibration. There are a number of important distinctions between screening and prone biopsy units. For the two currently available prone units, a pronounced heel effect makes ion chamber position critical. Focal spot measurements are particularly difficult on one unit because there is no light field. The fixed grid on the other unit must be tested with a flood film. Physicists who inspect these units before their clinical use should be aware of variations needed by this equipment for specific acceptance tests.
引用
收藏
页码:1197 / 1201
页数:5
相关论文
共 5 条
[1]  
BARNES GT, 1991, SCREEN FILM MAMMOGRA
[2]   STEREOTAXIC INSTRUMENT FOR NEEDLE-BIOPSY OF MAMMA [J].
BOLMGREN, J ;
JACOBSON, B ;
NORDENSTROM, B .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1977, 129 (01) :121-125
[3]   CALCULATION OF DOSE AND CONTRAST FOR 2 MAMMOGRAPHIC GRIDS [J].
DANCE, DR ;
PERSLIDEN, J ;
CARLSSON, GA .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :235-248
[4]  
HENDRICK RE, 1993, SYLLABUS CATEGORICAL, P233
[5]   MAMMOGRAPHY FOCAL SPOT MEASUREMENT WITH A STAR PATTERN - TECHNIQUES TO AVOID INACCURACIES [J].
KIMMESMITH, C ;
CHATZIIOANNOU, A .
MEDICAL PHYSICS, 1993, 20 (01) :93-97