DETERMINATION OF SURFACE DOSE-RATE FROM A SR-90 OPHTHALMIC APPLICATOR

被引:13
作者
ALI, MM
KHAN, FM
机构
[1] Department of Therapeutic Radiology—Radiation Oncology, University of Minnesota Hospital and Clinic, Minneapolis
关键词
D O I
10.1118/1.596480
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Surface dose rates from two 90Sr ophthalmic applicators were measured using thin thermoluminescent dosimeter (TLD) chips. The TLD’s were calibrated against the 90Sr source itself. The calibration technique is described. The results were compared to those obtained by three other techniques. The calibration in terms of roentgen equivalent betas per second provided by the manufacturer was found to be — 30% lower when compared to our measured dose rates. The surface dose rates obtained with the TLD’s calibrated against 6-MV x rays were 18% higher. One of the two sources was sent to both the National Institute of Standards and Technology (NIST) and Amersham International for calibration. NIST and Amersham use extrapolation chambers as a Bragg-Gray cavity to determine the surface dose rate. NIST results gave values higher than ours by —5%. The Amersham results were —35% lower than ours. © 1990, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:416 / 421
页数:6
相关论文
共 13 条
[1]  
[Anonymous], 1983, MED PHYS, V10, P741
[2]  
Berger M J., 1983, STOPPING POWERS RANG
[3]   POST-OPERATIVE IRRADIATION OF PTERYGIA - 10 MORE YEARS OF EXPERIENCE [J].
COOPER, JS .
RADIOLOGY, 1978, 128 (03) :753-756
[4]   THE POLARITY EFFECT FOR COMMERCIALLY AVAILABLE PLANE-PARALLEL IONIZATION CHAMBERS [J].
GERBI, BJ ;
KHAN, FM .
MEDICAL PHYSICS, 1987, 14 (02) :210-215
[5]   MEASUREMENT AND CORRECTION OF NONUNIFORM SURFACE DOSE RATES FROM BETA EYE APPLICATORS [J].
HENDEE, WR .
AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1968, 103 (04) :734-&
[6]   THERMOLUMINESCENT DOSIMETRY OF BETA DEPTH DOSE [J].
HENDEE, WR .
AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1966, 97 (04) :1045-&
[7]  
HOLT JG, 1979, INT J RADIAT ONCOL, V5, P2031, DOI 10.1016/0360-3016(79)90955-6
[8]   THERMOLUMINESCENT INTERNAL BETA-RAY DOSIMETRY [J].
KASTNER, J ;
HUKKOO, R ;
OLTMAN, BG ;
DAYAL, Y .
RADIATION RESEARCH, 1967, 32 (03) :625-&
[9]  
LEA DE, 1947, ACTIONS RAD LIVING C, P15
[10]   DESIGN AND OPERATION OF AN EXTRAPOLATION CHAMBER WITH REMOVABLE ELECTRODES [J].
LOEVINGE.R ;
TROTT, NG .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1966, 17 (02) :103-&