THE POINT-SPREAD FUNCTION OF X-RAY IMAGE-INTENSIFIERS CCD-CAMERA AND IMAGING-PLATE SYSTEMS IN CRYSTALLOGRAPHY - ASSESSMENT AND CONSEQUENCES FOR THE DYNAMIC-RANGE

被引:26
作者
BOURGEOIS, D
MOY, JP
SVENSSON, SO
KVICK, A
机构
[1] European Synchrotron Radiaton, Facility, Grenoble
关键词
D O I
10.1107/S0021889894004048
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The point-spread functions (PSFs) of two X-ray CCD detectors and two imaging-plate (IF) scanners were compared using the monochromatic beam of the Materials Science beamline of the ESRF. The CCD detectors were a commercial medical type X-ray image intensifier (XRII) for energies above 20 keV and a tube with a beryllium window especially designed for the 5-50 keV range, optically coupled to a high-resolution cooled charge-coupled-device (CCD) camera. The IP scanners were the Fuji BAS2000 and the Molecular Dynamics PhosphorImager 400E, with plates from Fuji and Kodak. The PSFs were recorded using a 30 x 30 mu m pinhole at energies from 8 to 41 keV under normal and highly saturated conditions. Special care was taken in the study of PSF wings. Dynamic-range capabilities in the presence of PSF bleeding were also measured using a standard X-ray generator. The Fuji and MD IP scanners give very similar PSFs when used with Fuji plates [circa 130 mu m full width at half-maximum (FWHM), 1100 mu m at 0.1% of peak maximum]. The Kodak plates showed a broader PSF than the Fuji plates, one which also increases with X-ray energy. The standard XRII/CCD gave the coarsest PSF (circa 4000 mu m at 0.1% of peak maximum), whereas the finest PSF was recorded with the beryllium XRII/CCD used in magnified mode (900 mu m at 0.1%). The best dynamic range was obtained with the XRII/CCDs; the lowest with the Fuji scanner.
引用
收藏
页码:868 / 877
页数:10
相关论文
共 23 条
[1]   DESIGN AND PERFORMANCE OF AN IMAGING PLATE SYSTEM FOR X-RAY-DIFFRACTION STUDY [J].
AMEMIYA, Y ;
MATSUSHITA, T ;
NAKAGAWA, A ;
SATOW, Y ;
MIYAHARA, J ;
CHIKAWA, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 266 (1-3) :645-653
[2]  
Amemiya Y., 1990, SYNCHROTRON RAD NEWS, V3, P21, DOI [10.1080/08940889008602552, DOI 10.1080/08940889008602552]
[3]  
ARNDT UW, 1988, ADV ELECTRON EL PHYS, V74, P285
[4]   FOURIER SPECTROSCOPY AND CAUSALITY PRINCIPLE [J].
BARTHOLD.E ;
ERNST, RR .
JOURNAL OF MAGNETIC RESONANCE, 1973, 11 (01) :9-19
[5]   PROTEIN CRYSTALLOGRAPHIC DATA ACQUISITION AND PRELIMINARY-ANALYSIS USING KODAK STORAGE PHOSPHOR PLATES [J].
BILDERBACK, D ;
MOFFAT, K ;
OWEN, J ;
RUBIN, B ;
SCHILDKAMP, W ;
SZEBENYI, D ;
TEMPLE, BS ;
VOLZ, K ;
WHITING, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 266 (1-3) :636-644
[6]   PERFORMANCE OF ELECTRON IMAGE CONVERTERS WITH YAG SINGLE-CRYSTAL SCREEN AND CCD SENSOR [J].
DABERKOW, I ;
HERRMANN, KH ;
LIU, LB ;
RAU, WD .
ULTRAMICROSCOPY, 1991, 38 (3-4) :215-223
[7]  
DRIARD B, 1971, ADV ELECTRON ELECTRO, V33, P1031
[8]   HIGH-SENSITIVITY LENS-COUPLED SLOW-SCAN CCD CAMERA FOR TRANSMISSION ELECTRON-MICROSCOPY [J].
FAN, GY ;
ELLISMAN, MH .
ULTRAMICROSCOPY, 1993, 52 (01) :21-29
[9]  
HAMMERSLEY AP, 1994, IN PRESS NUCL INSTRU
[10]  
HELLIWELL JR, 1992, MACROMOLECULAR CRYST, P291