A general three-dimensional planning program has been developed for hyperthermia treatments for cancer.1 In this study, the program is used to analyze the three-dimensional temperature distributions generated by interstitial ferromagnetic implants. An empirical power absorption formula developed by Haider2 for thermally self-regulating nickel-silicon ferromagnetic seeds has been used to calculate the seed power absorption as a function of seed temperature. By properly choosing the seed type (Curie point of seeds), and by using appropriate seed spacing, this heating modality can generate desirable three-dimensional temperature distributions for many different situations over a fairly large range of blood perfusion values. Detailed information regarding the choice of the Curie points of the seeds and the seed spacing for certain blood perfusions is also given as a quantitative guide for treatment planning.
引用
收藏
页码:989 / 997
页数:9
相关论文
共 34 条
[1]   USABLE FREQUENCIES IN HYPERTHERMIA WITH THERMAL SEEDS [J].
ATKINSON, WJ ;
BREZOVICH, IA ;
CHAKRABORTY, DP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (01) :70-75
[2]   HYPERTHERMIA OF PET ANIMAL TUMORS WITH SELF-REGULATING FERROMAGNETIC THERMOSEEDS [J].
BREZOVICH, IA ;
LILLY, MB ;
MEREDITH, RF ;
WEPPELMANN, B ;
HENDERSON, RA ;
BRAWNER, W ;
SALTER, MM .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1990, 6 (01) :117-130
[3]   TEMPERATURE DISTRIBUTIONS IN TUMOR-MODELS HEATED BY SELF-REGULATING NICKEL COPPER ALLOY THERMOSEEDS [J].
BREZOVICH, IA ;
ATKINSON, WJ ;
CHAKRABORTY, DP .
MEDICAL PHYSICS, 1984, 11 (02) :145-152
[4]  
CETAS TC, 1989, HYPERTHERMIA ONCOLOG, V1, P863
[5]  
CHEN G, 1988, CANCER CHEMOTH PHARM, V22, P303
[6]  
Chen Z. P., 1989, THESIS U ARIZONA TUC
[7]   ERRORS BETWEEN TWO-DIMENSIONAL AND 3-DIMENSIONAL THERMAL-MODEL PREDICTIONS OF HYPERTHERMIA TREATMENTS [J].
CHEN, ZP ;
MILLER, WH ;
ROEMER, RB ;
CETAS, TC .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1990, 6 (01) :175-191
[8]  
CHEN ZP, 1991, J HYPERTHERMIA, V1, P735
[9]   TREATMENT PLANNING FOR FERROMAGNETIC SEED HEATING [J].
CHIN, RB ;
STAUFFER, PR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (02) :431-439
[10]   MATHEMATICAL-MODEL OF HUMAN TEMPERATURE REGULATORY SYSTEM - TRANSIENT COLD-EXPOSURE RESPONSE [J].
GORDON, RG ;
ROEMER, RB ;
HORVATH, SM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1976, 23 (06) :434-444