SCANNING SLIT CHEST RADIOGRAPHY - A PRACTICAL AND EFFICIENT SCATTER CONTROL DESIGN

被引:12
作者
BARNES, GT [1 ]
WU, XZ [1 ]
SANDERS, PC [1 ]
机构
[1] UNIV ALABAMA,BIRMINGHAM,AL 35233
关键词
RADIOGRAPHY; DIGITAL; TECHNOLOGY; THORAX;
D O I
10.1148/radiology.190.2.8284410
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To design a chest radiography unit that eliminates scatter. MATERIALS AND METHODS: A scanning slit radiography unit that also incorporates a conventional grid was designed. The scanning slit geometry defined a 10-mm-wide beam at the image receptor. RESULTS: With a typical heavy-duty radiographic tube, the proposed system was capable of delivering an entrance skin exposure of up to 32 mu C/kg for a 1-second scan. Monte Carlo simulations indicated that the scatter-to-primary ratio (S/P) imaged with the scanning slit system alone was 0.125 for a 35 x 43-cm, 23-cm-thick Lucite phantom imaged at 140 kVp. When the system was coupled with 4:1, 8:1, and 12:1 grids, S/P decreased to 0.063, 0.055, and 0.054, respectively. CONCLUSION: A scanning slit system coupled with a conventional grid is a practical and highly efficient method of controlling scatter and will result in minimal soft-tissue and bone measurement errors in dual-energy chest radiography.
引用
收藏
页码:525 / 528
页数:4
相关论文
共 16 条
[1]   CLINICAL-EVALUATION OF UNSHARP MASKING AND SLIT SCANNING TECHNIQUES IN CHEST RADIOGRAPHY [J].
ARMSTRONG, JD ;
SORENSON, JA ;
NELSON, JA ;
TOCINO, I ;
LESTER, PD ;
JANES, JO ;
NIKLASON, LT ;
STANISH, W .
RADIOLOGY, 1983, 147 (02) :351-356
[2]  
Barnes G T, 1979, Med Phys, V6, P197, DOI 10.1118/1.594562
[3]   DIGITAL CHEST RADIOGRAPHY - PERFORMANCE EVALUATION OF A PROTOTYPE UNIT [J].
BARNES, GT ;
SONES, RA ;
TESIC, MM .
RADIOLOGY, 1985, 154 (03) :801-806
[4]   THE SCANNING GRID - A NOVEL AND EFFECTIVE BUCKY MOVEMENT [J].
BARNES, GT ;
MORELAND, RF ;
YESTER, MV ;
WITTEN, DM .
RADIOLOGY, 1980, 135 (03) :765-767
[5]   REDUCTION OF SCATTER IN DIAGNOSTIC RADIOLOGY BY MEANS OF A SCANNING MULTIPLE SLIT ASSEMBLY [J].
BARNES, GT ;
CLEARE, HM ;
BREZOVICH, IA .
RADIOLOGY, 1976, 120 (03) :691-694
[6]   CALCIFICATION IN PULMONARY NODULES - DETECTION WITH DUAL-ENERGY DIGITAL RADIOGRAPHY [J].
FRASER, RG ;
HICKEY, NM ;
NIKLASON, LT ;
SABBAGH, EA ;
LUNA, RF ;
ALEXANDER, CB ;
ROBINSON, CA ;
KATZENSTEIN, ALA ;
BARNES, GT .
RADIOLOGY, 1986, 160 (03) :595-601
[7]   GENERALIZED IMAGE COMBINATIONS IN DUAL KVP DIGITAL RADIOGRAPHY [J].
LEHMANN, LA ;
ALVAREZ, RE ;
MACOVSKI, A ;
BRODY, WR ;
PELC, NJ ;
RIEDERER, SJ ;
HALL, AL .
MEDICAL PHYSICS, 1981, 8 (05) :659-667
[8]   SIMULATED PULMONARY NODULES - DETECTION WITH DUAL-ENERGY DIGITAL VERSUS CONVENTIONAL RADIOGRAPHY [J].
NIKLASON, LT ;
HICKEY, NM ;
CHAKRABORTY, DP ;
SABBAGH, EA ;
YESTER, MV ;
FRASER, RG ;
BARNES, GT .
RADIOLOGY, 1986, 160 (03) :589-593
[9]   SCATTERED RADIATION IN CHEST RADIOGRAPHY [J].
NIKLASON, LT ;
SORENSON, JA ;
NELSON, JA .
MEDICAL PHYSICS, 1981, 8 (05) :677-681
[10]   EFFECTS OF SCATTERED RADIATION AND VEILING GLARE IN DUAL-ENERGY TISSUE BONE IMAGING - A THEORETICAL-ANALYSIS [J].
SHAW, CG ;
PLEWES, DB .
MEDICAL PHYSICS, 1987, 14 (06) :956-967