AN OBJECTIVE MRTD FOR DISCRETE INFRARED IMAGING-SYSTEMS

被引:9
作者
LETTINGTON, AH
HONG, QH
机构
[1] J.J. Thomson Phys. Lab., Reading Univ.
关键词
D O I
10.1088/0957-0233/4/10/013
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
An objective method has been developed for measuring the minimum resolvable temperature difference (MRTD) of infrared imaging systems employing focal plane detector arrays. The resulting MRTD is uniquely determined by applying discrete Fourier analysis, thus avoiding the previous multi-valued problem associated with the non-isoplanatic property of discrete devices. The derivation is based on sinusoidal wave input targets but a practical measuring procedure has been developed using conventional bar patterns; alternatively the MRTD may be calculated by measuring the MTF using the line spread function.
引用
收藏
页码:1106 / 1110
页数:5
相关论文
共 12 条
[2]  
DANN RJ, 1986, SPIE INFRARED TECH, V685, P123
[3]  
DEJONG AN, 1988, SPIE, V916, P127
[4]   A DISCRETE MODULATION TRANSFER-FUNCTION FOR FOCAL PLANE ARRAYS [J].
LETTINGTON, AH ;
HONG, QH .
INFRARED PHYSICS, 1993, 34 (01) :109-114
[5]   MEASUREMENT OF THE DISCRETE MODULATION TRANSFER-FUNCTION [J].
LETTINGTON, AH ;
HONG, QH .
JOURNAL OF MODERN OPTICS, 1993, 40 (02) :203-212
[6]  
LLOYD JM, 1975, THERMAL IMAGING SYST
[7]  
MCCRACKEN WL, 1988, SPIE, V892, P38
[8]  
Oppenheim Alan V, 1975, DIGIT SIGNAL PROCESS
[9]   STATIC PERFORMANCE-MODEL FOR THERMAL IMAGING-SYSTEMS [J].
RATCHES, JA .
OPTICAL ENGINEERING, 1976, 15 (06) :525-530
[10]  
RATCHES JA, 1975, 7043 US ARM EL COMM