SIMULTANEOUS INTRAOPERATIVE MEASUREMENT OF CARDIAC-OUTPUT BY THERMODILUTION AND TRANSTRACHEAL DOPPLER

被引:19
作者
SIEGEL, LC
FITZGERALD, DC
ENGSTROM, RH
机构
[1] Department of Anesthesia, H3580, Stanford Univ Med Ctr, Stanford
关键词
HEART; CARDIAC OUTPUT; MEASUREMENT TECHNIQUE; DOPPLER; NONINVASIVE; TRANSTRACHEAL;
D O I
10.1097/00000542-199104000-00008
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Intraoperative measurement of cardiac output with transtracheal Doppler (DOP) was compared with that measured by thermodilution (TD). Cardiac output was measured simultaneously with both methods in 17 adult patients. For 86 pairs of measurements, the average difference between the two techniques was -0.21.min-1. This bias had a standard deviation of 1.71.min-1. The average of the absolute value of the difference between measurements made with the two techniques was 1.31.min-1, with a standard deviation of 1.11.min-1. The limits of agreement were -3.6 to 3.11.min-1. Linear regression yielded the following equation: DOP = 0.62 TD + 1.54 1.min-1 (r = 0.63). To evaluate the ability of transtracheal DOP to trend changes in cardiac output, the changes in cardiac output at sequential time points were compared for the two techniques. The average difference in the changes in cardiac output measured by the two techniques was 0.01.min-1. This bias had a standard deviation of 1.7 1.min-1. In conclusion, the transtracheal DOP technique did not reproduce the measurement of cardiac output by TD intraoperatively. Transtracheal DOP did not accurately trend changes in the TD measurement. These findings were obtained from patients with cardiovascular disease, and the conclusions may depend in part on the patient population and the investigators' experience with the transtracheal DOP technique.
引用
收藏
页码:664 / 669
页数:6
相关论文
共 9 条
[1]   TRANS-TRACHEAL DOPPLER - A NEW PROCEDURE FOR CONTINUOUS CARDIAC-OUTPUT MEASUREMENT [J].
ABRAMS, JH ;
WEBER, RE ;
HOLMEN, KD .
ANESTHESIOLOGY, 1989, 70 (01) :134-138
[2]   CONTINUOUS CARDIAC-OUTPUT DETERMINATION USING TRANS-TRACHEAL DOPPLER - INITIAL RESULTS IN HUMANS [J].
ABRAMS, JH ;
WEBER, RE ;
HOLMEN, KD .
ANESTHESIOLOGY, 1989, 71 (01) :11-15
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   RELATION BETWEEN PRESSURE AND DIAMETER IN ASCENDING AORTA OF MAN [J].
GREENFIELD, JC ;
PATEL, DJ .
CIRCULATION RESEARCH, 1962, 10 (05) :778-+
[5]   COMPARING METHODS OF MEASUREMENT - AN ALTERNATIVE APPROACH [J].
LAMANTIA, KR ;
OCONNOR, T ;
BARASH, PG .
ANESTHESIOLOGY, 1990, 72 (05) :781-783
[6]   EFFECT OF INJECTATE VOLUME AND TEMPERATURE ON THERMODILUTION CARDIAC-OUTPUT DETERMINATION [J].
PEARL, RG ;
ROSENTHAL, MH ;
NIELSON, L ;
ASHTON, JPA ;
BROWN, BW .
ANESTHESIOLOGY, 1986, 64 (06) :798-801
[7]  
PIERPONT GL, 1988, CIRCULATION, V78, P351
[8]   AN EVALUATION OF THERMODILUTION CARDIAC-OUTPUT MEASUREMENT USING THE SWAN-GANZ CATHETER [J].
RUNCIMAN, WB ;
ILSLEY, AH ;
ROBERTS, JG .
ANAESTHESIA AND INTENSIVE CARE, 1981, 9 (03) :208-220
[9]   THERMODILUTION CARDIAC-OUTPUT MEASUREMENT - EFFECTS OF THE RESPIRATORY CYCLE ON ITS REPRODUCIBILITY [J].
STEVENS, JH ;
RAFFIN, TA ;
MIHM, FG ;
ROSENTHAL, MH ;
STETZ, CW .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1985, 253 (15) :2240-2242