PHOTOCOAGULATION THROUGH GOLDMANN CONTACT GLASS .4. A SECOND MODEL OF ATTACHMENT TO ZEISS-OBERKOCHEN PHOTOCOAGULATOR AND SOME IMPROVEMENTS OF SIEMENS RUBY LASER COAGULATOR

被引:17
作者
LOTMAR, W
FANKHAUSER, F
ROULIER, A
机构
[1] Swiss Office of Weights & Measures, Wabern
[2] Universitäts-Augenklinik, Bern
关键词
D O I
10.1001/archopht.1969.00990020316003
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A second prototype of the optical attachment to the Zeiss-Oberkochen photocoagulator allowing the use of the Goldmann contact glass is presented, showing improvements mainly on the observation system by the addition of a slit lamp and allowing for fluorescein funduscopy. Some development of the ruby-laser coagulator described in part II is also reported. A contact glass with a cover glass coated by an antireflection layer is proposed. © 1969, American Medical Association. All rights reserved.
引用
收藏
页码:314 / +
页数:1
相关论文
共 5 条
[1]  
FANKHAUSER F, 1967, ARCH OPHTHALMOL-CHIC, V77, P320
[2]   PHOTOCOAGULATION THROUGH GOLDMANN CONTACT GLASS .2. AN APPARATUS USING A QUASI-CONTINUOUS LASER SOURCE [J].
FANKHAUSER, F ;
LOTMAR, W ;
ROULIER, A .
ARCHIVES OF OPHTHALMOLOGY, 1968, 79 (06) :674-+
[3]   PHOTOCOAGULATION THROUGH GOLDMANN CONTACT GLASS .3. CLINICAL EXPERIENCE WITH AN APPARATUS USING A QUASI-CONTINUOUS LASER SOURCE [J].
FANKHAUSER, F ;
LOTMAR, W ;
ROULIER, A .
ARCHIVES OF OPHTHALMOLOGY, 1968, 79 (06) :684-+
[4]  
FANKHAUSER F, 1968, MOD PROBL OPHTHAL, V7, P256
[5]  
HAINING WM, 1968, ARCH OPHTHALMOL-CHIC, V79, P10