CONTINUOUS MEASUREMENT OF BLOOD-GASES INVIVO BY MASS-SPECTROMETRY

被引:6
作者
FOEX, P
BURT, G
MAYNARD, P
RYDER, WA
HAHN, CEW
机构
[1] Nuffield Department of Anaesthetics, University of Oxford
基金
英国医学研究理事会;
关键词
D O I
10.1093/bja/51.11.999
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Continuous analysis of oxygen and carbon dioxide tension in the blood phase, over periods of 4-5 h, was carried out in the pulmonary artery (93 determinations in six anaesthetized dogs) and in the aorta (29 determinations in four anaesthetized dogs). Silastic-covered stainless steel catheters attached to a mass spectrometer were used. The mass spectrometer signals were linearly related to the blood-gas tensions measured by conventional analysis. The mass spectrometer signals were calibrated in vivo by exposing the animals to high and low oxygen and carbon dioxide tension. With such in vivo calibration the slopes of the straight line regressions for mass spectrometer Po2 or Pco2 on Po2 or Pco2 by conventional analysis were comprised between 0.944 and 1.031 while the standard error of the slopes were between 0.019 and 0.031. Deliberate reductions of cardiac output had little effect on the mass spectrometer readings. © 1979 Macmillan Journals Ltd.
引用
收藏
页码:999 / 1005
页数:7
相关论文
共 27 条
[1]   A NONTHROMBOGENIC DIFFUSION MEMBRANE FOR CONTINUOUS IN-VIVO MEASUREMENT OF BLOOD GASES BY MASS SPECTROMETRY [J].
BRANTIGA.JW ;
GOTT, VL ;
VESTAL, ML ;
FERGUSSO.GJ ;
JOHNSTON, WH .
JOURNAL OF APPLIED PHYSIOLOGY, 1970, 28 (03) :375-&
[2]   TEFLON MEMBRANE FOR MEASUREMENT OF BLOOD AND INTRAMYOCARDIAL GAS TENSIONS BY MASS SPECTROSCOPY [J].
BRANTIGAN, JW ;
GOTT, VL ;
MARTZ, MN .
JOURNAL OF APPLIED PHYSIOLOGY, 1972, 32 (02) :276-+
[3]   CLINICAL CATHETER FOR CONTINUOUS BLOOD-GAS MEASUREMENT BY MASS-SPECTROMETRY [J].
BRANTIGAN, JW ;
DUNN, KL ;
ALBO, D .
JOURNAL OF APPLIED PHYSIOLOGY, 1976, 40 (03) :443-446
[4]  
DARDIK H, 1970, SURG GYNECOL OBSTETR, V131, P1157
[5]   RESPIRATORY MASS SPECTROMETER [J].
FOWLER, KT .
PHYSICS IN MEDICINE AND BIOLOGY, 1969, 14 (02) :185-&
[7]   MEASUREMENT OF OXYGEN MICROCATHODE CURRENTS BY MEANS OF A FIELD-EFFECT TRANSISTOR OPERATIONAL AMPLIFIER SYSTEM WITH DIGITAL DISPLAY [J].
HAHN, CEW .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1969, 2 (01) :48-&
[8]  
HASS WK, 1968, CIRCULATION, V38, P96
[9]   A MASS SPECTROMETER FOR CONTINUOUS GAS ANALYSIS [J].
HUNTER, JA ;
STACY, RW ;
HITCHCOCK, FA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1949, 20 (05) :333-336
[10]   NOMOGRAMS FOR CORRECTION OF BLOOD PO2, PCO2, PH AND BASE EXCESS FOR TIME AND TEMPERATURE [J].
KELMAN, GR ;
NUNN, JF .
JOURNAL OF APPLIED PHYSIOLOGY, 1966, 21 (05) :1484-&