FATAL PULMONARY HEMORRHAGE AFTER USE OF THE FLOW-DIRECTED BALLOON-TIPPED CATHETER

被引:92
作者
PAPE, LA
HAFFAJEE, CI
MARKIS, JE
OCKENE, IS
PARASKOS, JA
DALEN, JE
ALPERT, JS
机构
[1] PETER BENT BRIGHAM HOSP, DIV CARDIOVASC, BOSTON, MA 02115 USA
[2] BETH ISRAEL HOSP, DIV CARDIOVASC, BOSTON, MA 02215 USA
[3] HARVARD UNIV, SCH MED, BOSTON, MA 02115 USA
关键词
D O I
10.7326/0003-4819-90-3-344
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The flow-directed balloon-tipped catheter is extensively used in intensive care units, catheterization laboratories, operating rooms, and emergency wards. Major complications associated with its use have not been frequently reported. In a recent 2-year period in four hospitals, the authors identified five cases of fatal pulmonary hemorrhage resulting from balloon-tipped catheters. The authors review here four additional cases previously cited in the literature and discuss possible mechanisms and predisposing factors associated with this complication and guidelines for safe use of these catheters. Pulmonary artery rupture is probably commoner than previously reported.
引用
收藏
页码:344 / 347
页数:4
相关论文
共 15 条
[1]   COMPLETE HEART-BLOCK CAUSED BY SWAN-GANZ CATHETER [J].
ABERNATHY, WS .
CHEST, 1974, 65 (03) :349-349
[2]  
BRAUNWAL.E, 1968, CIRCULATION, V37, pIII1
[3]  
CHUN GMH, 1971, NEW ENGL J MED, V284, P1041
[4]   PULMONARY COMPLICATIONS OF FLOW-DIRECTED BALLOON-TIPPED CATHETER [J].
FOOTE, GA ;
SCHABEL, SI ;
HODGES, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1974, 290 (17) :927-931
[5]   FATAL PULMONARY HEMORRHAGE COMPLICATING USE OF A FLOW-DIRECTED BALLOON-TIPPED CATHETER IN A PATIENT RECEIVING ANTICOAGULANT THERAPY [J].
GOLDEN, MS ;
PINDER, T ;
ANDERSON, WT ;
CHEITLIN, MD .
AMERICAN JOURNAL OF CARDIOLOGY, 1973, 32 (06) :865-867
[6]   HEMOPTYSIS WITH FLOW-DIRECTED CARDIAC CATHETERIZATION [J].
LAPIN, ES ;
MURRAY, JA .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1972, 220 (09) :1246-&
[7]  
LEMEN R, 1975, NEW ENGL J MED, V292, P211
[8]  
LIPP H, 1971, NEW ENGL J MED, V284, P220
[9]   POTENTIAL ARTIFACT IN MEASUREMENT OF LEFT-VENTRICULAR FILLING PRESSURE WITH FLOW-DIRECTED CATHETERS [J].
MEISTER, SG ;
ENGEL, TR ;
FISCHER, HA ;
FRANKL, WS .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1976, 2 (02) :175-179
[10]  
MEISTER SG, 1971, CATHET CARDIOVASC DI, V3, P171