SOME EFFECTS OF LOW ACID CONCENTRATIONS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY

被引:51
作者
MARICHY, M
MERMET, M
MERMET, JM
机构
[1] Laboratoire des Sciences Analytiques, Université Claude Bernard - Lyon I, 69622 Villeurbanne Cedex
关键词
D O I
10.1016/0584-8547(90)80063-O
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The effect of HCl, HNO3, and HClO4, was studied at concentrations below 1% ( v v) in inductively coupled plasma atomic emission spectrometry. An increase in the signal was obtained for HCl with a maximum intensity corresponding to a concentration near 0.001%, in contrast to the depression effect obtained with HClO4. No significant effect was observed for HNO3. These phenomena cannot be explained by a change in nebuliser or transport efficiency. No variation in the aerosol droplet size distribution was found. The only major change observed was the decrease (50%) in the add concentration of the aerosol compared to that of the solution, irrespective of the type of acid or the use of a surfactant. © 1990.
引用
收藏
页码:1195 / 1201
页数:7
相关论文
共 26 条
[1]   REMOVAL OF A MATRIX EFFECT IN ICP-AES MULTIELEMENT ANALYSIS BY SIMPLEX OPTIMIZATION [J].
BELCHAMBER, RM ;
BETTERIDGE, D ;
WADE, AP ;
CRUICKSHANK, AJ ;
DAVISON, P .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (05) :503-505
[2]   INTERFERENCE EFFECTS FROM AEROSOL IONIC REDISTRIBUTION IN ANALYTICAL ATOMIC SPECTROMETRY [J].
BOROWIEC, JA ;
BOORN, AW ;
DILLARD, JH ;
CRESSER, MS ;
BROWNER, RF ;
MATTESON, MJ .
ANALYTICAL CHEMISTRY, 1980, 52 (07) :1054-1059
[4]  
CANALS A, IN PRESS J ANAL ATOM
[5]  
CARRE M, IN PRESS
[6]  
Church S. E., 1981, DEV ATOMIC PLASMA SP, P410
[7]   INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION SPECTROMETRY - ANALYSIS OF BIOLOGICAL-MATERIALS AND SOILS FOR MAJOR, TRACE, AND ULTRA-TRACE ELEMENTS [J].
DAHLQUIST, RL ;
KNOLL, JW .
APPLIED SPECTROSCOPY, 1978, 32 (01) :1-29
[8]   POSSIBILITIES FOR CORRECTING THE INFLUENCE OF MINERAL ACIDS AND OF DRIFT IN ICP EMISSION SPECTRAL-ANALYSIS [J].
DELIJSKA, A ;
VOUCHKOV, M .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1985, 321 (05) :448-452
[9]   INFLUENCE OF SOLUTION UPTAKE RATE ON SIGNALS AND INTERFERENCES IN INDUCTIVELY COUPLED PLASMA OPTICAL-EMISSION SPECTROMETRY [J].
FARINO, J ;
MILLER, JR ;
SMITH, DD ;
BROWNER, RF .
ANALYTICAL CHEMISTRY, 1987, 59 (18) :2303-2309
[10]   CONTRIBUTION OF THE CHARGE-TRANSFER PROCESS TO THE EXCITATION MECHANISMS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROSCOPY [J].
GOLDWASSER, A ;
MERMET, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (07) :725-739