IMPROVING AGREEMENT BETWEEN RADIATION-DELINEATED FIELD EDGES ON SIMULATION AND PORTAL FILMS - THE EDGE TOLERANCE-TEST TOOL

被引:6
作者
MCCULLOUGH, EC [1 ]
MCCOLLOUGH, KP [1 ]
SIMON, SP [1 ]
机构
[1] GOOD SAMARITAN HOSP,DEPT RADIAT ONCOL,PHOENIX,AZ 85006
关键词
THERAPEUTIC RADIOLOGY; QUALITY ASSURANCE; PORTAL FILM; TEST TOOL; TREATMENT MACHINE; SIMULATOR;
D O I
10.1118/1.597129
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The anatomy at the edge of a treatment machine portal film may differ from that shown by the delineator lines on a simulator film by up to 10 mm, even if both the simulator and the treatment machine meet accepted criteria for mechanical tolerances. To assure that this possibility is minimized requires some form of overall alignment check between the simulator and the treatment machine. A new test device, the edge tolerance test tool (ET3), has been designed to permit a quick and accurate check on whether portal film disagreement with simulator films is due to an accumulation of tolerances. Its use should eliminate one source of this common problem.
引用
收藏
页码:375 / 376
页数:2
相关论文
共 1 条
[1]  
1984, AAPM13 AM ASS PHYS M