ENERGY IMPARTED TO NEONATES DURING X-RAY EXAMINATIONS IN A SPECIAL CARE BABY UNIT

被引:37
作者
CHAPPLE, CL [1 ]
FAULKNER, K [1 ]
HUNTER, EW [1 ]
机构
[1] ROYAL VICTORIA INFIRM,DEPT RADIOL,NEWCASTLE TYNE NE1 4LP,TYNE & WEAR,ENGLAND
关键词
D O I
10.1259/0007-1285-67-796-366
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Neonates in a special care baby unit may receive a large number of X-rays and their dosimetry is of particular importance. A method of calculating energy imparted to neonates has been developed and a survey carried out in one unit, over a period of 18 months. Entrance dose was calculated from the technique factors used and measurement of tube output. Technique factors were recorded by the radiographer for each exposure taken, and output was measured both with and without an incubator present. Field size was determined by measurements made retrospectively from the radiograph, and a Monte Carlo simulation was used to determine factors for conversion to imparted energy. 119 neonates were included in the survey, and the mean total energy imparted was found to be 0.09 mJ. The maximum value was a factor of nine greater than this. The study also highlighted the potential for dose reduction with regard to better collimation and shielding.
引用
收藏
页码:366 / 370
页数:5
相关论文
共 20 条
[1]   A STANDARD TIMING BENCHMARK FOR EGS4 MONTE-CARLO CALCULATIONS [J].
BIELAJEW, AF ;
ROGERS, DWO .
MEDICAL PHYSICS, 1992, 19 (02) :303-304
[2]   COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR [J].
BIRCH, R ;
MARSHALL, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) :505-517
[3]   AN INVESTIGATION INTO THE PERFORMANCE OF AN AUTOMATED QUALITY ASSURANCE AND DOSIMETRY SYSTEM IN DIAGNOSTIC-RADIOLOGY [J].
CHAPPLE, CL ;
FAULKNER, K ;
HARRISON, RM .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (752) :635-639
[4]   RADIATION-DOSE TO NEONATES ON A SPECIAL CARE BABY UNIT [J].
FAULKNER, K ;
BARRY, JL ;
SMALLEY, P .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (735) :230-233
[5]   THE RISK OF DIAGNOSTIC RADIATION OF THE NEWBORN [J].
FLETCHER, EWL ;
BAUM, JD ;
DRAPER, G .
BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (698) :165-170
[6]  
GREENING JR, 1981, FUNDAMENTALS RAD DOS, P75
[7]   BACKSCATTER FACTORS FOR DIAGNOSTIC-RADIOLOGY (1-4 MM AL HVL) [J].
HARRISON, RM .
PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (12) :1465-1474
[8]  
Jones D. G., 1985, NRPBR186 NAT RAD PRO
[9]  
MURRAY D, 1990, Australasian Physical and Engineering Sciences in Medicine, V13, P132
[10]  
PERSLIDEN J, 1993, ACTA RADIOL, V34, P92