PROXIMAL FEMORAL FRACTURES - A BIOMECHANICAL STUDY TO COMPARE INTRAMEDULLARY AND EXTRAMEDULLARY FIXATION

被引:115
作者
CURTIS, MJ
JINNAH, RH
WILSON, V
CUNNINGHAM, BW
机构
[1] The Johns Hopkins University, Baltimore, MD
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 1994年 / 25卷 / 02期
关键词
D O I
10.1016/0020-1383(94)90111-2
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
This study assesses the rigidity and strength of fixation presided by intramedullary and extramedullary devices for proximal femoral fractures. Stable and unstable intertrochanteric fractures were studied in paired femora after internal fixation with the Gamma nail and Richards 135 degrees classic hip-screw implants; in subtrochanteric fractures, the 95 degrees Richards condylar screw was studied in addition. Subsidence of the sliding screw within the plate and nail constructs was measured. Fixation of subtrochanteric fractures with the intramedullary Gamma nail was significantly stronger and more rigid than that with the extramedullary screw plate devices. Under conditions of simulated protected weight bearing, the 95 degrees condylar screw plate provided more rigid fixation than did the 135 degrees hip screw. There was no significant difference in the strength of fixation of stable and unstable intertrochanteric fractures between the Gamma nail and the hip screw, although the Gamma nail provided more rigid fixation.
引用
收藏
页码:99 / 104
页数:6
相关论文
共 34 条
[1]  
Apel, Patwardhan, Pinzur, Dobozi, Axial loading studies of unstable intertrochanteric fractures of the femur., Clin Orthop Relat Res, 246, (1989)
[2]  
Bonamo, Accettola, Treatment of intertrochanteric fractures with a sliding nail-plate, J. Trauma, 22, (1982)
[3]  
Bridle, Patel, Bircher, Calvert, Fixation of intertrochanteric fractures of the femur, J. Bone Joint Surg. [Br], 73 B, (1991)
[4]  
British Standards Institute, Draft for development. Method for determination of endurance properties of stemmed femoral components of hip joint prostheses (DD91:1986), (1986)
[5]  
Calvert, The Gamma nail — a significant advance or a passing fashion?, J. Bone Joint Surg [Br], 74 B, (1992)
[6]  
Christie, Court-Brown, Kinninmonth, Howie, Intramedullary locking nails in the management of femoral shaft fractures, J. Bone Joint Surg. [Br], 70 B, (1988)
[7]  
Clark, Crofts, Saleh, Femoral neck fracture fixation. Comparison of a sliding screw with lag screws, J. Bone Joint Surg. [Br], 72 B, (1990)
[8]  
Crowninshield, Brand, Johnston, The effects of walking velocity and age on hip kinematics and kinetics, Clin. Orthop., 132, (1978)
[9]  
Davis, Sher, Horsman, Simpson, Porter, Checketts, Intertrochanteric femoral fractures. Mechanical failure after internal fixation, J. Bone Joint Surg. [Br], 72 B, (1990)
[10]  
Halder, The Gamma nail for peritrochanteric fractures, J. Bone Joint Surg. [Br], 74 B, (1992)