VASCULAR COMPLICATIONS AFTER BALLOON AND NEW DEVICE ANGIOPLASTY

被引:216
作者
POPMA, JJ [1 ]
SATLER, LF [1 ]
PICHARD, AD [1 ]
KENT, KM [1 ]
CAMPBELL, A [1 ]
CHUANG, YC [1 ]
CLARK, C [1 ]
MERRITT, AJ [1 ]
BUCHER, TA [1 ]
LEON, MB [1 ]
机构
[1] WASHINGTON HOSP CTR, DEPT INTERNAL MED, DIV CARDIOL, WASHINGTON, DC 20010 USA
关键词
ANGIOPLASTY; PSEUDOANEURYSM; FISTULAS;
D O I
10.1161/01.CIR.88.4.1569
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Despite their potential advantages, new coronary angioplasty devices may be associated with more frequent vascular complications than noted after standard balloon angioplasty, theoretically due to the larger sheaths and prolonged periods of anticoagulation required by some of these devices. This study sought to identify the incidence, predictors, and clinical outcome of vascular complications after new device angioplasty. Methods and Results. The clinical course of 1413 patients was reviewed after balloon or new device angioplasty. Vascular complications were defined as formation of a pseudoaneurysm, arteriovenous fistula, retroperitoneal hematoma, or groin hematoma associated with a > 15 -point hematocrit drop or the need for surgical repair. Stepwise logistic regression was used to identify independent predictors for vascular complications. Vascular complications developed after 84 (5.9%) procedures; they occurred more frequently after intracoronary stenting (14.0%) and extraction atherectomy (12.5%) than after balloon angioplasty (3.2%) (odds ratios, 4.86; P<.001, and 4.26, P<.05, respectively). Independent predictors of vascular complications included the use of intraprocedural thrombolytic agents (P<.01), intracoronary stenting (P<.005), or extraction atherectomy (P<.05); high maximum creatinine level (P<.005); low nadir platelet count (P<.001); longer periods of excess anticoagulation (P<.05); and the need for repeat coronary angioplasty (P<.005). Vascular complications were not related to the size of the arterial sheath used. Conclusions. Vascular complications developed more frequently after new device angioplasty than after balloon angioplasty, with the risk for vascular complications directly related to the degree of periprocedural anticoagulation.
引用
收藏
页码:1569 / 1578
页数:10
相关论文
共 28 条
[1]  
BAIM DS, 1985, CARDIAC CATHETERIZAT, P59
[2]  
Dixon W. J., 1990, BMDP STATISTICAL SOF
[3]   PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - REPORT OF COMPLICATIONS FROM THE NATIONAL-HEART-LUNG-AND-BLOOD-INSTITUTE PTCA REGISTRY [J].
DORROS, G ;
COWLEY, MJ ;
SIMPSON, J ;
BENTIVOGLIO, LG ;
BLOCK, PC ;
BOURASSA, M ;
DETRE, K ;
GOSSELIN, AJ ;
GRUNTZIG, AR ;
KELSEY, SF ;
KENT, KM ;
MOCK, MB ;
MULLIN, SM ;
MYLER, RK ;
PASSAMANI, ER ;
STERTZER, SH ;
WILLIAMS, DO .
CIRCULATION, 1983, 67 (04) :723-730
[4]  
GLASER RL, 1989, ARCH SURG-CHICAGO, V124, P1313
[5]   THE DISTINCTION BETWEEN FEMORAL-ARTERY PSEUDOANEURYSMS AND OTHER CAUSES OF GROIN MASSES - VALUE OF DUPLEX DOPPLER SONOGRAPHY [J].
HELVIE, MA ;
RUBIN, JM ;
SILVER, TM ;
KRESOWIK, TF .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 150 (05) :1177-1180
[6]  
HOLMES D R, 1992, Journal of the American College of Cardiology, V19, p76A
[7]  
Hosmer DW, 1989, APPLIED LOGISTIC REG
[8]   PERIPHERAL VASCULAR COMPLICATIONS FROM PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - A COMPARISON WITH TRANS-FEMORAL CARDIAC-CATHETERIZATION [J].
KAUFMAN, J ;
MOGLIA, R ;
LACY, C ;
DINERSTEIN, C ;
MOREYRA, A .
AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 1989, 297 (01) :22-25
[9]   ROLE OF NEW TECHNOLOGY IN BALLOON ANGIOPLASTY [J].
KING, SB .
CIRCULATION, 1991, 84 (06) :2574-2579
[10]   A PROSPECTIVE-STUDY OF THE INCIDENCE AND NATURAL-HISTORY OF FEMORAL VASCULAR COMPLICATIONS AFTER PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY [J].
KRESOWIK, TF ;
KHOURY, MD ;
MILLER, BV ;
WINNIFORD, MD ;
SHAMMA, AR ;
SHARP, WJ ;
BLECHA, MB ;
CORSON, JD .
JOURNAL OF VASCULAR SURGERY, 1991, 13 (02) :328-335