EFFECT OF THE THERMODYNAMICS OF AN INFANT PLETHYSMOGRAPH ON THE MEASUREMENT OF THORACIC GAS VOLUME

被引:16
作者
SLY, PD [1 ]
LANTERI, C [1 ]
BATES, JHT [1 ]
机构
[1] MCGILL UNIV,MEAKINS CHRISTIE LABS,MONTREAL H3A 2T5,QUEBEC,CANADA
关键词
digital correction; Thermal time constant; effects of frequency; fan in box; copperwool in box;
D O I
10.1002/ppul.1950080312
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Adult plethysmographs have frequency responses that are essentially flat over the range of frequencies encountered in the measurement of thoracic gas volume (TGV). An infant plethysmograph is necessarily much smaller than an adult model. This means that there is a smaller mean distance over which heat diffusion must occur between the air in the plethysmograph and its walls. This in turn leads to a much reduced thermal time constant. We examined the effects of thermal time constant of a 60 L infant plethysmograph on measurements of TGV in infants. The thermal time constant was measured by rapidly injecting 20 mL of air into the plethysmograph, and found to be 0.16 ± 0.09 s when the box was empty. We calculated from this time constant that measurements of TGV should be quite dependent on the frequency at which the associated panting maneuvers are performed. TGV was measured in 5 infants less than 6 months old in the recovery phase following acute viral bronchiolitis. When we performed a digital correction of the measurements, to compensate for the thermal time constant of the plethysmograph, the TGV values decreased by a mean of 12%. Agitating the air in the plethysmograph with a fan decreased the thermal time constant of the box and reduced measured TGV by a mean of 8.4%. These results indicate that thermodynamics of infant plethysmographs can be an important source of error in TGV measurements. Pediatr Pulmonol 1990; 8:203‐208. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:203 / 208
页数:6
相关论文
共 9 条
[1]   MECHANICS OF RESPIRATION IN UNANESTHETIZED GUINEA PIGS [J].
AMDUR, MO ;
MEAD, J .
AMERICAN JOURNAL OF PHYSIOLOGY, 1958, 192 (02) :364-368
[2]  
BARGETON DM, 1969, BODY PLETHYSMOGRAPHY, V4, P2
[3]   PROBLEMS IN MEASUREMENT OF THORACIC GAS VOLUME IN INFANCY [J].
BEARDSMORE, CS ;
STOCKS, J ;
SILVERMAN, M .
JOURNAL OF APPLIED PHYSIOLOGY, 1982, 52 (04) :995-999
[4]  
BRACEWELL R, 1978, FOURIER TRANSFORM IT
[5]   RAPID PLETHYSMOGRAPHIC METHOD FOR MEASURING THORACIC GAS VOLUME - COMPARISON WITH A NITROGEN WASHOUT METHOD FOR MEASURING FUNCTIONAL RESIDUAL CAPACITY IN NORMAL SUBJECTS [J].
DUBOIS, AB ;
BOTELHO, SY ;
BEDELL, GN ;
MARSHALL, R ;
COMROE, JH .
JOURNAL OF CLINICAL INVESTIGATION, 1956, 35 (03) :322-326
[7]   RESPONSE OF NORMAL INFANTS TO INHALED HISTAMINE [J].
LESOUEF, PN ;
GEELHOED, GC ;
TURNER, DJ ;
MORGAN, SEG ;
LANDAU, LI .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (01) :62-66
[8]   INHERENT VARIABILITY OF PULMONARY-FUNCTION TESTS IN INFANTS WITH BRONCHIOLITIS [J].
MALLOL, J ;
HIBBERT, ME ;
ROBERTSON, CF ;
OLINSKY, A ;
PHELAN, PD ;
SLY, PD .
PEDIATRIC PULMONOLOGY, 1988, 5 (03) :152-157
[9]  
PESLIN R, 1984, TECHNIQUES RESPIRA 2