PACEMAKER FAILURE INDUCED BY RADIOTHERAPY

被引:60
作者
SOULIMAN, SK [1 ]
CHRISTIE, J [1 ]
机构
[1] WESTERN INFIRM & ASSOCIATED HOSP,DEPT CARDIOL,DEPT CLIN PHYS & BIOENGN,GLASGOW,SCOTLAND
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 03期
关键词
PACEMAKER; RADIATION THERAPY; MULTIPROGRAMMABLE; RADIATION DAMAGE;
D O I
10.1111/j.1540-8159.1994.tb01387.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
It is well documented that radiation therapy can have significant transitory and permanent effects on the function of cardiac pacemakers. In this study 18 multiprogrammable pacemakers were tested to establish any pattern by which these pacemakers are affected. The results confirm previous predictions on how pacemakers are damaged. The damage could be divided essentially into three types: (1) temporary change to interference or safety mode pacing lasting for the duration of the irradiation only; (2) change to interference mode pacing-from which recovery may occur after reprogramming the pacemaker; and (3) severe damage; in this case the pacemaker stops generating pulses. Recovery from this may occur a long time after the end of the treatment, but this recovery is mostly incomplete and the pacemaker cannot be used reliably thereafter. Therefore it is essential that patients with implanted pacemakers undergoing radiation therapy, should be monitored closely during the course of the treatment and for a few weeks thereafter.
引用
收藏
页码:270 / 273
页数:4
相关论文
共 12 条
[1]   DAMAGING EFFECT OF THERAPEUTIC RADIATION ON PROGRAMMABLE PACEMAKERS [J].
ADAMEC, R ;
HAEFLIGER, JM ;
KILLISCH, JP ;
NIEDERER, J ;
JAQUET, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :146-150
[2]   X-RAY-EFFECTS ON PACEMAKER TYPE CIRCUITS [J].
BLAMIRES, NG ;
MYATT, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :151-155
[3]   PACEMAKER FAILURE ASSOCIATED WITH THERAPEUTIC RADIATION [J].
BROOKS, C ;
MUTTER, M .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1988, 6 (06) :591-593
[4]   PACEMAKER FAILURE DUE TO RADIATION-THERAPY [J].
KATZENBERG, CA ;
MARCUS, FI ;
HEUSINKVELD, RS ;
MAMMANA, RB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1982, 5 (02) :156-159
[5]   RADIATION-INDUCED FAILURES OF COMPLEMENTARY METAL-OXIDE SEMICONDUCTOR CONTAINING PACEMAKERS - A POTENTIALLY LETHAL COMPLICATION [J].
LEWIN, AA ;
SERAGO, CF ;
SCHWADE, JG ;
ABITBOL, AA ;
MARGOLIS, SC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (10) :1967-1969
[6]   RADIATION TESTING COMPLEMENTARY (SYMMETRY) METAL-OXIDE SEMICONDUCTOR (CMOS) ARRAYS FOR SATELLITES [J].
MATTEUCCI, AJ ;
SCHNEIDER, MF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (06) :2285-2288
[7]   THE EFFECT OF THERAPEUTIC X-RADIATION ON A SAMPLE OF PACEMAKER GENERATORS [J].
MAXTED, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (09) :1143-1146
[8]   RADIATION EFFECT ON IMPLANTED PACEMAKERS [J].
POURHAMIDI, AH .
CHEST, 1983, 84 (04) :499-500
[9]   PACEMAKER FAILURE RESULTING FROM RADIATION-DAMAGE [J].
QUERTERMOUS, T ;
MEGAHY, MS ;
DASGUPTA, DS ;
GRIEM, ML .
RADIOLOGY, 1983, 148 (01) :257-258
[10]  
SALMI J, 1990, STRAHLENTHER ONKOL, V166, P153