TRANSCATHETER EMBOLIZATION COIL CLOSURE OF PATENT DUCTUS-ARTERIOSUS - MODIFIED DELIVERY FOR ENHANCED CONTROL DURING COIL POSITIONING
被引:26
作者:
KUHN, MA
[1
]
LATSON, LA
[1
]
机构:
[1] CLEVELAND CLIN FDN,CTR PEDIAT & CONGENITAL HEART DIS,CLEVELAND,OH 44121
来源:
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS
|
1995年
/
36卷
/
03期
关键词:
INTERVENTIONAL CARDIOLOGY;
CONGENITAL HEART DISEASE;
PATENT DUCTUS ARTERIOSUS;
D O I:
10.1002/ccd.1810360325
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
The use of Gianturco coils (Cook, Inc., Bloomington, IN) has been recently described as a method of occluding restrictive patent ductus arteriosus (PDA). Precise control of the coil during positioning or withdrawal is difficult. We designed an inexpensive and readily available system to control the coil during delivery and repositioning. A 5 French Mallincrodt vertebral catheter was heat tapered to allow slight, but definite resistance of a 0.035'' guide wire when it was passed through the catheter. The delivery system has been used to successfully coil occlude restrictive PDAs in seven patients. The system has added no additional costs to the procedure and has provided improved control of the coil position prior to release of the coil. (C) 1995 Wiley-Liss, Inc.