SAR PATTERNS OF EXTERNAL 915 MHZ MICROWAVE APPLICATORS

被引:32
作者
STRAUBE, WL
MYERSON, RJ
EMAMI, B
LEYBOVICH, LB
机构
[1] Radiation Oncology Center, Hyperthermia Section, Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, MO
关键词
Applicators; Hyperthermia; Quality assurance;
D O I
10.3109/02656739009140962
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Analysis of the results from recent clinical trials has shown that tumour size is a significant prognostic factor in eventual tumour control in patients treated with thermoradiotherapy. The critical issue appears to be the adequate coverage of hyperthermia target volume with therapeutic temperature. Therefore one must choose appropriate applicators for the treatment of a given tumour. Accurate knowledge of performance characteristics of the applicators used in clinics thus becomes crucial. In an attempt to take the first step for the appropriate selection of applicators in clinics several commonly used applicators were evaluated according to their 75% 50% and 25% two-dimensional SAR (specific absorption rate) contours at depths of 1-3 cm. The data were subsequently approximated by rectangles. This type of information, even with its limitations, is extremely important in implementing quality assurance in hyperthermia. In this communication we will present such information, and the implications in current hyperthermia clinical trials will be discussed. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:665 / 670
页数:6
相关论文
共 5 条
[1]   A PRACTICAL, MODULAR HYPERTHERMIA PHANTOM [J].
LEYBOVICH, LB ;
NUSSBAUM, GH .
MEDICAL PHYSICS, 1984, 11 (02) :207-208
[2]  
LEYBOVICH LB, 1989, UNPUB INT J HYPERTHE
[3]  
MYERSON RJ, 1988, UNPUB INT J RAD ONCO
[4]  
PEREZ CA, 1988, IN PRESS INT J RAD O
[5]   IMPORTANT PROGNOSTIC FACTORS INFLUENCING OUTCOME OF COMBINED RADIATION AND HYPERTHERMIA [J].
VALDAGNI, R ;
LIU, FF ;
KAPP, DS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 15 (04) :959-972