CONTROLLED EVALUATION OF BACTEC PLUS 26 AND ROCHE SEPTI-CHEK AEROBIC BLOOD CULTURE BOTTLES

被引:31
作者
WEINSTEIN, MP
MIRRETT, S
WILSON, ML
HARRELL, LJ
STRATTON, CW
RELLER, LB
机构
[1] UNIV MED & DENT NEW JERSEY,ROBERT WOOD JOHNSON MED SCH,DEPT MED,NEW BRUNSWICK,NJ 08901
[2] UNIV MED & DENT NEW JERSEY,ROBERT WOOD JOHNSON MED SCH,DEPT PATHOL,NEW BRUNSWICK,NJ 08901
[3] DUKE UNIV,MED CTR,CLIN MICROBIOL LAB,DURHAM,NC 27710
[4] DUKE UNIV,SCH MED,DEPT PATHOL,DURHAM,NC 27710
[5] DUKE UNIV,SCH MED,DEPT MICROBIOL & IMMUNOL,DURHAM,NC 27710
[6] DUKE UNIV,SCH MED,DEPT MED,DURHAM,NC 27710
[7] VANDERBILT UNIV,MED CTR,SCH MED,MICROBIOL LAB,NASHVILLE,TN 37232
[8] VANDERBILT UNIV,SCH MED,DEPT PATHOL,NASHVILLE,TN 37232
[9] VANDERBILT UNIV,SCH MED,DEPT MED,NASHVILLE,TN 37232
关键词
D O I
10.1128/JCM.29.5.879-882.1991
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Blood culture bottles that accept up to 10 ml of blood have recently been introduced for the BACTEC nonradiometric blood culture system. The new formulation, designated BACTEC Plus, contains 25 ml of tryptic soy broth, 0.05% sodium polyanetholesulfonate, and mixed resins. In a collaborative study conducted at three university hospitals, we evaluated the BACTEC Plus 26 (BP26) aerobic bottle and the Roche Septi-Chek aerobic bottle with its agar slide paddle in 5,293 paired blood cultures. Significantly more microorganisms (P < 0.001), especially Staphylococcus aureus (P < 0.001), Staphylococcus epidermidis (P < 0.01), enterococci (P < 0.005), and members of the family Enterobacteriaceae (P < 0.005), were detected by the BP26 bottle. When both bottles detected growth, BP26 did so earlier (P < 0.001). In particular, S. epidermidis (P < 0.001), streptococci (P < 0.005), enterococci (P < 0.05), and members of the family Enterobacteriaceae (P < 0.001) were detected earlier by the BP26 bottle. We conclude that the BP26 bottle provides a yield and speed of detection of microorganisms superior to those of the Roche Septi-Chek aerobic blood culture bottle.
引用
收藏
页码:879 / 882
页数:4
相关论文
共 25 条
[1]  
APPLEBAUM PC, 1983, J CLIN MICROBIOL, V17, P48
[2]   EVALUATION OF THE ANTIBIOTIC REMOVAL DEVICE [J].
APPLEMAN, MD ;
SWINNEY, RS ;
HESELTINE, PNR .
JOURNAL OF CLINICAL MICROBIOLOGY, 1982, 15 (02) :278-281
[3]   BLOOD CULTURES [J].
ARONSON, MD ;
BOR, DH .
ANNALS OF INTERNAL MEDICINE, 1987, 106 (02) :246-253
[4]   ROUTINE EVALUATION OF BACTEC NR-16A AND NR-17A MEDIA [J].
COURCOL, RJ ;
DUROCHER, AV ;
ROUSSELDELVALLEZ, M ;
FRUCHART, A ;
MARTIN, GR .
JOURNAL OF CLINICAL MICROBIOLOGY, 1988, 26 (09) :1619-1622
[5]   DETECTION OF BACTEREMIA IN PATIENTS RECEIVING ANTI-MICROBIAL THERAPY - AN EVALUATION OF THE ANTI-MICROBIAL REMOVAL DEVICE AND B-16 MEDIUM [J].
DOERN, GV ;
GANTZ, NM .
JOURNAL OF CLINICAL MICROBIOLOGY, 1983, 18 (01) :43-48
[6]   MICROBIOLOGICAL AND CLINICAL-EVALUATION OF THE ISOLATOR LYSIS-CENTRIFUGATION BLOOD CULTURE TUBE [J].
HENRY, NK ;
MCLIMANS, CA ;
WRIGHT, AJ ;
THOMPSON, RL ;
WILSON, WR ;
WASHINGTON, JA .
JOURNAL OF CLINICAL MICROBIOLOGY, 1983, 17 (05) :864-869
[7]   ROUTINE USE OF BACTEC B-16 BOTTLES TO REMOVE ANTI-BACTERIAL AND ANTI-TUMOR AGENTS FROM BLOOD CULTURES OF CANCER-PATIENTS [J].
HOPFER, RL ;
MOORE, DG ;
FAINSTEIN, V ;
WATKINS, S ;
WENGLAR, M .
JOURNAL OF CLINICAL MICROBIOLOGY, 1983, 18 (04) :759-764
[8]   THE IMPORTANCE OF VOLUME OF BLOOD CULTURED IN THE DETECTION OF BACTEREMIA AND FUNGEMIA [J].
ILSTRUP, DM ;
WASHINGTON, JA .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1983, 1 (02) :107-110
[9]  
JUNGKIND DL, 1989, 1989 ANN M AM SOC MI, P431
[10]  
Lennette E.H., 1980, MANUAL CLIN MICROBIO, P195