A NEW FLUORESCENT X-RAY SOURCE FOR PHOTON SCATTERING INVESTIGATIONS

被引:19
作者
HARDING, G
JORDAN, B
KOSANETZKY, J
机构
[1] PHILIPS MED SYST,W-2000 HAMBURG 63,GERMANY
[2] PHILIPS GMBH IND X-RAY SYST,W-2000 HAMBURG 63,GERMANY
关键词
D O I
10.1088/0031-9155/36/12/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A description is given of the design and performance of a novel fluorescent x-ray source, comprising a demountable conical target enclosed in a conical anode x-ray tube. Measurements are presented of the spectral purity of the fluorescent emission and the variation in photon flux with the applied tube voltage. A Ta target tube (K(alpha-1) = 57.532 keV) powered by a standard 3.75 kW, 160 kV high voltage generator has a source brightness of 10(9) photons/s sr mm2. ThiS is 500 times greater than that of the brightest available Am-241 radionuclide sources. Some potential applications of the fluorescent x-ray source are listed.
引用
收藏
页码:1573 / 1583
页数:11
相关论文
共 15 条
[2]   X-RAY WAVELENGTHS [J].
BEARDEN, JA .
REVIEWS OF MODERN PHYSICS, 1967, 39 (01) :78-&
[3]  
Birch R., 1979, SCI REPORT SERIES, V30
[4]   PHOTON SCATTERING IN BIOMEDICALLY IMPORTANT ELEMENTS [J].
BRADLEY, DA ;
GHOSE, AM .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (11) :1385-1397
[5]   PHOTOELECTRON BREMSSTRAHLUNG - ANALYTICAL POSSIBILITIES [J].
CHETTLE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (02) :259-264
[6]  
Drexler G., 1975, Proceedings of the symposium on advances in biomedical dosimetry, P499
[7]   CHARACTERISTIC X-RAYS - STILL DEVELOPING SUBJECT [J].
DYSON, NA .
PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (01) :1-29
[8]  
HARDING G, 1986, SPIE, V671, P81
[9]   ASSESSMENT OF GAMMA-RAY SCATTERING FOR THE CHARACTERIZATION OF BIOLOGICAL-MATERIAL [J].
HOLT, RS ;
KOURIS, K ;
COOPER, MJ ;
JACKSON, DF .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (12) :1435-1440
[10]  
Hubbell J. H., 1975, Journal of Physical and Chemical Reference Data, V4, P471, DOI 10.1063/1.555523