HISTOLOGICAL FEATURES OF THE INTERFACE MEMBRANE OF FAILED ISOELASTIC CEMENTLESS PROSTHESES

被引:1
作者
BOSS, JH
SHAJRAWI, I
SOUDRY, M
MENDES, DG
机构
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The interface membranes of both components from two isoelastic cementless hip arthroplasties, which were revised for aseptic loosening, were examined histologically. The membranes consisted of densely structured fibrous tissue interspersed with giant cell granulomas, sheets of histiocytes and chronic inflammatory infiltrates. The granulomas and the histiocytes contained polymeric and metallic wear products, respectively. The morphological features of these membranes were compared with those of cemented joint replacements. Apart from the deposits of acrylic cement bordered by macrophages, there were no distinctive differences between the membranes of failed cementless and cemented arthroplasties. The release and deposition of wear products, of whatever nature, are apparently responsible for the formation of exuberant interface membranes.
引用
收藏
页码:399 / 403
页数:5
相关论文
empty
未找到相关数据