THROMBIN-SOAKED EMBOLIZATION COILS - THE EFFECT ON WHOLE-BLOOD CLOTTING TIME

被引:3
作者
NICHOLSON, DA
COCKBURN, JF
BRADSHAW, AE
DAWSON, P
机构
[1] HAMMERSMITH HOSP,ROYAL POSTGRAD MED SCH,DEPT DIAGNOST RADIOL,LONDON W12 0HS,ENGLAND
[2] HAMMERSMITH HOSP,ROYAL POSTGRAD MED SCH,DEPT HAEMATOL,LONDON W12 0HS,ENGLAND
关键词
D O I
10.1016/S0009-9260(05)80313-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Embolization coils are well established as embolic agents for the treatment of various conditions. Several authors have commented on the increased 'thrombogenicity' of coils following soaking in thrombin solutions. We have carried out an in vitro study, carefully measuring the effect on whole blood clotting time (WBCT), of soaking coils in thrombin solutions of different concentrations (100, 200, 400, 1000 U/ml). Untreated steel coils are shown to have clot promoting activity (CPA) in vitro, reducing WBCT from 14.85 min to 5.53 min. Passing the coils down a saline-filled catheter slightly reduces their CPA, but not significantly (p = 0.21). With thrombin concentrations above 100 U/ml, a significant reduction in WBCT is recorded, but although there is a trend of increasing CPA with increasing thrombin concentration from 200-1000 U/ml, a plateau in WBCT is seen, and the difference is not significant. It therefore appears that the clot promoting activity of embolic coils is significantly increased by soaking them in a relatively weak thrombin solution. The use of such a solution (e.g. 200 U/ml) in vivo would have obvious value in limiting the potential systemic effects of thrombin.
引用
收藏
页码:108 / 110
页数:3
相关论文
共 11 条
[1]   PERCUTANEOUS LIVER-BIOPSY AND TRACK EMBOLIZATION WITH STEEL COILS [J].
ALLISON, DJ ;
ADAM, A .
RADIOLOGY, 1988, 169 (01) :261-263
[2]   CHRONIC VASCULAR REACTIONS TO STEEL COIL OCCLUSION DEVICES [J].
BARTH, KH ;
STRANDBERG, JD ;
KAUFMAN, SL ;
WHITE, RI .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1978, 131 (03) :455-458
[3]   COAGULATION OF ANEURYSMS BY DIRECT PERCUTANEOUS THROMBIN INJECTION [J].
COPE, C ;
ZEIT, R .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1986, 147 (02) :383-387
[4]  
DEBRUN G, 1981, AM J NEURORADIOL, V2, P167
[5]   EMBOLIZATION OF A SUBCLAVIAN ARTERY ANEURYSM WITH STEEL COILS AND THROMBIN [J].
FEDULLO, LM ;
MERANZE, SG ;
MCLEAN, GK ;
BURKE, DR .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1987, 10 (03) :134-137
[6]   MECHANICAL DEVICES FOR ARTERIAL-OCCLUSION [J].
GIANTURCO, C ;
ANDERSON, JH ;
WALLACE, S .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1975, 124 (03) :428-435
[7]  
Harman J T, 1991, J Vasc Interv Radiol, V2, P166, DOI 10.1016/S1051-0443(91)72493-7
[8]  
KERBER C, 1975, INVEST RADIOL, V10, P272
[9]  
LAMMERT GK, 1989, AJR AM J ROENTGENOL, V152, P383
[10]   STEEL OCCLUSION COILS - PRETREATMENT WITH THROMBIN [J].
MCLEAN, GK ;
STEIN, EJ ;
BURKE, DR ;
MERANZE, SG .
RADIOLOGY, 1986, 158 (02) :549-550