ERRORS IN FLUID BALANCE WITH PUMP CONTROL OF CONTINUOUS HEMODIALYSIS

被引:23
作者
ROBERTS, M
WINNEY, RJ
机构
[1] ROYAL INFIRM,MED RENAL UNIT,EDINBURGH EH3 9YW,MIDLOTHIAN,SCOTLAND
[2] VET ADM MED CTR,DEPT NEPHROL,SEPULVEDA,CA 91343
关键词
HEMODIALYSIS; PUMPS; CONTINUOUS HEMODIALYSIS; PUMP CONTROL; CAVHD; CVVHD;
D O I
10.1177/039139889201500207
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of pumps both proximal and distal to the dialyzer during continuous hemodialysis provides control of dialysate and ultrafiltration flow rates, thereby reducing nursing time. However, we had noted unexpected severe extracellular fluid depletion suggesting that errors in pump delivery may be responsible. We measured in vitro the operation of various pumps under conditions similar to continuous hemodialysis. Fluid delivery of peristaltic and roller pumps varied with how the tubing set was inserted in the pump. Piston and peristaltic pumps with dedicated pump segments were more accurate. Pumps should be calibrated and tested under conditions simulating continuous hemodialysis prior to in vivo use.
引用
收藏
页码:99 / 102
页数:4